5. Refer to the
provided with V.A.C. Dressings
for information on using excess V.A.C. Drape material to seal the leak area.
6. Once the leak is resolved using the Seal Check Tool, press
on the Seal Check
screen to return to the ‘Home’ screen.
Dressing Application Instructions for Use
®
®
™
™
Exit
WARNING: Under clinician supervision, replace V.A.C. Dressing with alternate
dressing if therapy is interrupted for more than two hours.
®
If alarm condition cannot be resolved, contact KCI. See back cover of this User Manual
for country specific contact information.
This alarm screen appears when the ActiV.A.C. Therapy Unit
has detected a leak that has not been resolved and therapy
has been interrupted.
This alarm will be accompanied by a repeating audible tone.
This alarm may be silenced for two (2) minutes during
troubleshooting by pressing
.
®
‘Audio Pause’
To resolve this alarm:
1. Press ‘
’ on this screen to return to the ‘
.
Reset
Home Screen’
If the leak condition was not resolved, the ActiV.A.C. Therapy Unit will take several minutes to re-
identify that a leak exists before the
will reappear and provide the opportunity to
continue troubleshooting a leak using the
Tool.
The patient’s only access to the Seal Check Tool is through the Leak Alarm screen.
Should the leak not be resolved within three minutes, this screen will once again appear and
therapy will stop.
®
™
Leak Alarm
Seal Check
™
Audio Pause
Reset
Leak Alarm
Therapy Interrupted
WARNING:
Press ? For more information
Leak Alarm Therapy Interrupted
Alerts and Alarms (cont.)
On/Off
2. Restart therapy by pressing
.
4.
While therapy is on and using light pressure, move your hand and fingers slowly around the
edges of the drape and the SensaT.R.A.C. Pad. The bar graph will lower and the audible tone (if
is on) will slow down when the leak is found.
®
Seal Audio
7. Once the leak is resolved, ensure that V.A.C. Therapy is on by observing that the green
crescent is lit on the
button and the icon is rotating on the ‘Home’ screen.
®
On/Off
If this alarm is
resolved within 3 minutes, the
alarm
will appear and therapy will stop.
not
Leak Alarm Therapy Interrupted
Leak Alarm (cont.)