KBR multimess D9-PQ
20144 EDEBD
A0228-5023-1_EN
16
Electrical safety
– IEC 61010-1
– IEC 61010-2-030
Protection class
1
Pollution degree
2
Overvoltage category
mains supply option :
H1
H2
300V / CAT III
150V / CAT III
High voltage test
Pulse voltage 6 kV
5 sec 5,4 kV RMS
1 min 3,6kV RMS
Measurement category 300V / CAT IV
600V / CAT III
Altitude
≤ 2000m
Environmental parameters
Storage and transport
Operation
Ambient temperature :
Limit range of operation
IEC 60721-3-1 / 1K5
-40 ... +70°C
IEC 60721-3-2 / 2K4
-40 ... +70°C
IEC 60721-3-3 / 3K6
-25 ... +55°C
Ambient temperature :
Rated range of operation
---
IEC 60721-3-3 / 3K5
mod. -10 ... +45°C
Relative humidity: 24h average
No condensation or ice
5...95 %
5...95 %
Solar radiations
---
700W/m2
Vibration, earth tremors
IEC 60721-3-1 / 1M1
IEC 60721-3-2 / 2M1
IEC 60721-3-3 / 3M1