Instructions for use For SONICflex tips paro - 0.571.0371, paro A - 1.006.2020
6 Preparation methods according to ISO 17664 | 6.1 Preparation at the site of use

12 / 18

6 Preparation methods according to ISO 17664

 WARNING

Hazard from non-sterile products.

Risk of infection to the care provider and patient.

▶ Before first use and after each use, prepare and sterilise the medical device.

Note

The following preparation procedures apply for the SONICflex tips, torque spanner

and nozzle needle.

6.1 Preparation at the site of use

 WARNING

Hazard from nonsterile products.

There is a risk of infection from contaminated medical devices.

▶ Take suitable personal protective measures.

Note

Do not place the tips in the drill bit bath, as the fine capillaries would corrode badly,

making it impossible to rinse them under running water.

Note

If a sterile coolant is used, the SONICflex tips must be rinsed with spray water after

each use to prevent crystallisation in the tips.
▶ Remove all residual cement, composite or blood without delay.

▶ The medical device must be dry when transported for reconditioning.

▶ Do not place it in a solution or similar.

▶ Reprocess the medical device as soon as possible after treatment.

▶ Prepare the SONICflex instrument according to the manufacturer's instructions for

use.

6.2 Preparation before cleaning

▶ Remove the tip from the SONICflex using the torque wrench.

6.3 Cleaning

 CAUTION

Malfunctions from cleaning in the ultrasonic unit.

Defects in the product.

▶ Only clean manually or in a thermodisinfector.

6.3.1 Manual cleaning of the exterior

Accessories required:

▪ Tap water 30 ℃ +/- 5 ℃ (86 

o

F +/- 10 

o

F)

▪ Nozzle needle

Содержание SONICflex tips paro

Страница 1: ...Instructions for use For SONICflex tips paro 0 571 0371 paro A 1 006 2020 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...2 Instructions for use 10 6 Preparation methods according to ISO 17664 12 6 1 Preparation at the site of use 12 6 2 Preparation before cleaning 12 6 3 Cleaning 12 6 3 1 Manual cleaning of the exterior...

Страница 4: ...afely Symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Can be steam sterilised at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Thermodisinfe...

Страница 5: ...indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER...

Страница 6: ...ng on them gently with your thumb or forefinger In addition expose the tips to approx 10 N 1 kg mechanical load without func tion CAUTION Hazard from worn SONICflex paro tips SONICflex paro tips can b...

Страница 7: ...with KaVo SONICflex LUX 2000 L LX N NX 2000 NM LM 2004 LM 2003 2003 L xx Tip identification by 2 digit number Tips with short thread for use with KaVo SONICflex quick 2008 xxA Tip identification by 2...

Страница 8: ...Instructions for use For SONICflex tips paro 0 571 0371 paro A 1 006 2020 3 Product description 3 3 Transportation and storage conditions 8 18 Protect from moisture...

Страница 9: ...in the torque wrench This may result in a risk of injury for the user When inserting the tip into the torque spanner make sure that the tip s end al ways faces into the opening of the torque spanner...

Страница 10: ...nd premature breakage of the tip Level 3 should only be used briefly 5 2 Instructions for use This tip set can be used for initial treatment and regular recall Regular recall treat ments are very impo...

Страница 11: ...y to identify The SONICflex paro tips are designed for use in the following areas 60 60 A SONICflex paro straight no 60 60 A all quadrants 61 61 A SONICflex paro left angled no 61 61 A mandible left v...

Страница 12: ...bit bath as the fine capillaries would corrode badly making it impossible to rinse them under running water Note If a sterile coolant is used the SONICflex tips must be rinsed with spray water after...

Страница 13: ...ith tip screwed on into the adapter for interior cleaning in the thermodisinfector e g Miele accessories for dental silicon adapters The program settings and cleansers and disinfectants that must be u...

Страница 14: ...ONICflex instru ment KaVo recommends thermodisinfectors in accordance with ISO 15883 1 e g Miele G 7781 G 7881 Validation was performed with the program VARIO TD the cleaner neodisher mediclean the ne...

Страница 15: ...product Immediately remove the product from the steam steriliser after the sterilisation cy cle Autoclave with a triple prevacuum for at least 4 minutes at 134 o C 1 273o F 1 8 Autoclave using the gr...

Страница 16: ...0 571 0371 paro A 1 006 2020 7 Tools 6 8 Storage 16 18 7 Tools Obtainable from the dentalmed specialist supplier Material summary Mat no Torque wrench 1 000 4887 Nozzle needle 0 410 0911 Sterilisatio...

Страница 17: ......

Страница 18: ...1 007 1938 kb 20090227 01 US en...

Отзывы: