Instructions for use For SONICflex tips paro - 0.571.0371, paro A - 1.006.2020

3 Product description | 3.1

3 Product description

SONICflex par tip set – REF 0.571.0371
SONICflex paro tip set A – REF 1.006.2020

3.1

3.2 Technical data

① Tip no. 60 straight – REF 0.571.7402
① Tip no. 60 A straight – REF 1.006.1934
② Tip no. 61 left – REF 0.571.7412
② Tip no. 61 A left – REF 1.006.1935
③ Tip no. 62 right – REF 0.571.7422
③ Tip no. 62 A right – REF 1.006.1936

3.2.1 Identification tip type

Tips with long thread for use with:

KaVo SONICflex LUX 2000 L / LX / N / NX, 2000 NM / LM, 2004 LM, 2003 / 2003 L

xx

Tip identification by:
① 2-digit number

Tips with short thread for use with:

KaVo SONICflex quick 2008

xxA

Tip identification by:
② 2-digit number and capital letter A

3.3 Transportation and storage conditions

 CAUTION

It is hazardous to start up the medical device after it has been stored strongly

refrigerated.

This can cause the medical device to malfunction.

▶ Prior to start-up, very cold products must be heated to a temperature of 20

o

C to

25

o

C (68

o

F to 77

o

F).

Temperature: -20°C to +70°C (-4°F to +158°F)

Relative humidity: 5% RH to 95% RH absence of condensation

Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)

7 / 18

Содержание SONICflex tips paro

Страница 1: ...Instructions for use For SONICflex tips paro 0 571 0371 paro A 1 006 2020 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...2 Instructions for use 10 6 Preparation methods according to ISO 17664 12 6 1 Preparation at the site of use 12 6 2 Preparation before cleaning 12 6 3 Cleaning 12 6 3 1 Manual cleaning of the exterior...

Страница 4: ...afely Symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Can be steam sterilised at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Thermodisinfe...

Страница 5: ...indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER...

Страница 6: ...ng on them gently with your thumb or forefinger In addition expose the tips to approx 10 N 1 kg mechanical load without func tion CAUTION Hazard from worn SONICflex paro tips SONICflex paro tips can b...

Страница 7: ...with KaVo SONICflex LUX 2000 L LX N NX 2000 NM LM 2004 LM 2003 2003 L xx Tip identification by 2 digit number Tips with short thread for use with KaVo SONICflex quick 2008 xxA Tip identification by 2...

Страница 8: ...Instructions for use For SONICflex tips paro 0 571 0371 paro A 1 006 2020 3 Product description 3 3 Transportation and storage conditions 8 18 Protect from moisture...

Страница 9: ...in the torque wrench This may result in a risk of injury for the user When inserting the tip into the torque spanner make sure that the tip s end al ways faces into the opening of the torque spanner...

Страница 10: ...nd premature breakage of the tip Level 3 should only be used briefly 5 2 Instructions for use This tip set can be used for initial treatment and regular recall Regular recall treat ments are very impo...

Страница 11: ...y to identify The SONICflex paro tips are designed for use in the following areas 60 60 A SONICflex paro straight no 60 60 A all quadrants 61 61 A SONICflex paro left angled no 61 61 A mandible left v...

Страница 12: ...bit bath as the fine capillaries would corrode badly making it impossible to rinse them under running water Note If a sterile coolant is used the SONICflex tips must be rinsed with spray water after...

Страница 13: ...ith tip screwed on into the adapter for interior cleaning in the thermodisinfector e g Miele accessories for dental silicon adapters The program settings and cleansers and disinfectants that must be u...

Страница 14: ...ONICflex instru ment KaVo recommends thermodisinfectors in accordance with ISO 15883 1 e g Miele G 7781 G 7881 Validation was performed with the program VARIO TD the cleaner neodisher mediclean the ne...

Страница 15: ...product Immediately remove the product from the steam steriliser after the sterilisation cy cle Autoclave with a triple prevacuum for at least 4 minutes at 134 o C 1 273o F 1 8 Autoclave using the gr...

Страница 16: ...0 571 0371 paro A 1 006 2020 7 Tools 6 8 Storage 16 18 7 Tools Obtainable from the dentalmed specialist supplier Material summary Mat no Torque wrench 1 000 4887 Nozzle needle 0 410 0911 Sterilisatio...

Страница 17: ......

Страница 18: ...1 007 1938 kb 20090227 01 US en...

Отзывы: