Instructions for use For SONICflex tips paro - 0.571.0371, paro A - 1.006.2020
6 Preparation methods according to ISO 17664 | 6.4 Disinfection
▪ Rubber eraser
▪ Brush, e.g. medium-hard toothbrush
Brush the SONICflex tip under running tap water using, e.g., a toothbrush of medium
hardness. Clean the cone of the tip with a rubber eraser. If necessary, clear the water
passage at the tip with a nozzle needle.
6.3.2 Mechanical cleaning of the exterior and interior
Note
Before setting up, screw the SONICflex tips clockwise onto the SONICflex instru‐
ment.
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
e.g. Miele G 7781 / G 7881.
(Validation was performed with the program "VARIO-TD", the cleaner "neodisher®
mediclean", the neutraliser "neodisher® Z" and rinse "neodisher® mielclear" and only
refers to the material compatibility with KaVo products.)
Note
Insert the instrument with tip screwed on into the adapter for interior cleaning in the
thermodisinfector (e.g. Miele accessories for dental silicon adapters).
▶ The program settings and cleansers and disinfectants that must be used can be
found in the instructions for use of the thermodisinfector.
6.3.3 Manual cleaning of the interior
To effectively set up, the inside of the machine must be cleaned automatically in a
cleaning and disinfection unit in accordance with ISO 15883-1.
(The interior of this product is not to be cleaned manually).
6.3.4
6.4 Disinfection
CAUTION
Malfunctioning from using a disinfectant bath or disinfectant containing chlor‐
ine.
Defects in the product.
▶ Only disinfect in a thermodisinfector or manually.
6.4.1 Manual disinfection of the the exterior
KaVo recommends the following products based on material compatibility. The micro‐
biological efficacy must be ensured by the disinfectant manufacturer.
▶ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▶ FD 322 made by Dürr
▶ CaviCide made by Metrex
Consumables required:
Cloths for wiping off the medical device.
13 / 18