Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805

3 Product description | 3.4 Scope of delivery

Electrical supply

Line voltage

100 to 240 VAC

Mains frequency

50 or 60 Hz

Power consumption

41 VA

Overvoltage category

II

Mains voltage fluctuation

±10 %

Compressed Air

Compressed air

0.4 to 0.6 MPa

Compressed air consumption

Approximately 50 NL/min

Air requirements

Dry, oil free, dirt free, uncontaminated in
accordance with EN ISO 7494-2

Operating environment

Temperature

15 to 40

o

C (59 to 104

o

F)

Humidity

25 to 90 %

Degree of soiling

2

Permissible to a maximum

2000 m above sea level (6560 feet alti‐
tude)

Transportation and storage conditions

Storage temperature

0 to 25

o

C (32 to 77

o

F)

Transport temperature

-18 to 40

o

C (-0 to 104

o

F)

Rel. humidity

5 to 95 % non-condensing

Air pressure

700 to 1060 hPa

3.4 Scope of delivery

Note

The package includes the articles listed below. If any of these articles is missing,

please contact the supplier immediately so that the missing article can be delivered.

15 / 36

Содержание QUATTROcare PLUS 2124 A

Страница 1: ...Instructions for use QUATTROcare PLUS 2124 A 1 008 3805 Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...2 3 2 Product specific 12 3 Product description 13 3 1 QUATTROcare PLUS 2124 A 13 3 2 Rating plate 14 3 3 Technical data 14 3 4 Scope of delivery 15 4 First use 17 4 1 Select location 17 4 2 Connect...

Страница 4: ...6 2 Safety checks STK 30 7 Troubleshooting 31 8 Tools and consumables 32 9 Terms and conditions of warranty 34 Instructions for use QUATTROcare PLUS 2124 A 1 008 3805 Table of contents 4 36...

Страница 5: ...tions AI Assembly instructions TI Technician s instructions SC Safety checks IEC International Electrotechnical Commission RI Repair instructions RK Retrofitting kit AS Assembly set EP Enclosed parts...

Страница 6: ...nt or recycling companies Com ply with the comprehensive return system KaVo has had its sales packaging licensed for this purpose Please comply with the regional public waste disposal system 1 3 2 Dam...

Страница 7: ...the shipping company Report the damage to KaVo Consult with KaVo first before returning a damaged product Send the signed delivery receipt to KaVo If the product is damaged but there was no discernab...

Страница 8: ...008 3805 1 User instructions 1 3 Transportation and storage 8 36 Humidity 5 to 95 non condensing Air pressure 700 to 1 060 hPa Transport upright Protect from impact Protect from moisture 15 kg max Per...

Страница 9: ...hazard levels are used in this docu ment to prevent personal and property damage CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNIN...

Страница 10: ...egal regulations must be observed during the use of the de vice in particular the following Applicable regulations governing the connection and start up of medical devices Current occupational safety...

Страница 11: ...unit and is spe cifically designed to fit this unit only The lubricant spray can is unable to be used manually Note Compliance with the preceding and subsequent re processing steps according to the re...

Страница 12: ...e device from heating up to more than 50 122o F CAUTION Contaminated compressed air Damage to and malfunction of the instrument In general ensure dry clean oil free compressed air in accordance with E...

Страница 13: ...ATTROcare PLUS 2124 A 3x INTRAmatic service couplings 1x MULTIflex service couplings Control panel and display Door of spray can chamber Chuck service adaptor QUATTROcare Plus Spray canister Collectin...

Страница 14: ...ATTROcare PLUS 2124 A SN Year of manufacture serial number REF Material number Fuse protected with T1000 mA Comply with the Instructions for use CE mark CSA mark Labelling according to 2002 96 EC 3 3...

Страница 15: ...ed in accordance with EN ISO 7494 2 Operating environment Temperature 15 to 40 o C 59 to 104 o F Humidity 25 to 90 Degree of soiling 2 Permissible to a maximum 2000 m above sea level 6560 feet alti tu...

Страница 16: ...of delivery 16 36 Chuck service adaptor QUATTROcare Plus Spray canister Compressed air coupling Instructions for use O ring MULTIflex 4 units O ring INTRAmatic 3 units Compressed air hose Power cable...

Страница 17: ...e the unit right next to a source of heat Keep away from ignition sources and do not use open fire close by Ensure sufficient ventilation of the premises CAUTION Danger of overheating Damage to and ma...

Страница 18: ...installation location Do not expose to di rect sunlight Do not set up the unit in cabinets Maintain a minimum distance of 50 cm 20 inches from electrical plugs light switches and sources of ignition...

Страница 19: ...Spray canister WARNING Use of servicing products other than original KaVo products The use of third party products can shorten the product life cause explosion damage or malfunction Always service KaV...

Страница 20: ...sheet of the spray can Note The safety data sheet of the QUATTROcare Plus Spray can be viewed at www kavo com Open the side door of the unit Swing the adaptor on the retention clip outward Place the...

Страница 21: ...lly and by touch for absence of leakage The spray can must rest flat against the adaptor There must be no gap between the spray can and the adaptor The connection must be tightly sealed Note If any oi...

Страница 22: ...t of the device Place the spray can on the retention clip and unscrew it counterclockwise from the adaptor Close the side door of the unit 4 3 3 Dispose of the spray canister Dispose of the spray can...

Страница 23: ...itch on off LED display Start button for chuck service Start button for service procedure 5 2 Open the front flap Note Objects on or in the front door Placing objects on or in the front door can damag...

Страница 24: ...ced because their necks are too thin These heads must be serviced together with the corresponding shanks Mount turbines on the MULTIflex service coupling service coupling on the far right and lock the...

Страница 25: ...he treatment Note The LED of the spray canister control lighting up indicates that the spray can is empty When the main switch is on the LED lights up Press the start key The automatic servicing proce...

Страница 26: ...in the axial direction of the service coupling and take it off Press on the latch with a thumb and take the straight or contra angle handpiece off by simultaneously pulling in the direction of the axi...

Страница 27: ...nd fasten it turning the union nut coun terclockwise against the direction of the arrow 5 7 Chuck Servicing Note Handpieces must be taken off the service couplings before the chuck service can be star...

Страница 28: ...e chuck to be serviced leading against the tip of the chuck service coupling Note If this is done with insufficient pressure there is a risk of the chuck service adaptor being pushed off the service c...

Страница 29: ...ntra angle handpieces and turbines can be changed Hold down the start key Turn the main switch on LEDs 1 to 4 of the LED display light up one after the other The blow out time is as follows for LED 1...

Страница 30: ...f the unit using a soft cloth and mild liquid cleanser Remove and clean the collecting pan The collecting pan can be washed in the dishwasher and in the thermodisinfector Temperature max 95 C 203 F Ti...

Страница 31: ...is ac tivated Insert original KaVospray canister Notify the KaVo Service Team Flashing LED light is on Bar code is not detected or cannot be read Insert original KaVospray canister Notify the KaVo Ser...

Страница 32: ...005 1707 Service coupling BORDEN 2 hole 1 002 1217 O ring INTRAmatic 0 200 6089 O ring MULTIflex 0 200 6098 O ring MULTIflex 0 200 6177 Seal for service coupling for INTRAmatic MULTIflex or heads 0 55...

Страница 33: ...8858 Service coupling MORITA Turbines 1 000 6063 Service coupling BIEN AIR 1 001 7860 Service coupling W H 1 001 7888 Presentation Material summary Mat No QUATTROcare Plus Spray canister 2140 P pack...

Страница 34: ...ith respect to compensation are excluded In the event of default and gross negligence or intent this shall only ap ply in the absence of mandatory legal regulations to the contrary KaVo cannot be held...

Страница 35: ......

Страница 36: ...1 008 6955 kb 20150122 13 en...

Отзывы: