Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805
2 Safety | 2.3 Safety instructions
▶ Please complete the disposal order entering the necessary data and submit it as
an online order or by Fax +49 (0) 3304 3919-590.
▶ The following contact options are also available for questions and for initiating a
disposal order:
Phone +49 (0) 3304 3919-500
enretec GmbH, Geschäftsbereich eomRECYCLING®, Kanalstraße 17, 16727
Velten, Germany
A unit that is not permanently installed in the office will be picked up.
A permanently installed unit will be picked up at the curb at your address on the
agreed date.
The owner or user of the device will have to bear the costs of disassembly, transporta‐
tion and packaging.
2.2.2 Product-specific
The intended use of the KaVo QUATTROcare is for internal cleaning, i.e, purging of
old lubricant, for the maintenance of rotating dental and surgical instruments. KaVo
QUATTROcare spray can is a component of the KaVo QUATTROcare unit and is spe‐
cifically designed to fit this unit only. The lubricant spray can is unable to be used
manually.
Note
Compliance with the preceding and subsequent re-processing steps according to
the recommendations of the Medical Devices Law and RKI guidelines is required al‐
though the QUATTROcare PLUS 2124 A services the straight and contra-angle
handpieces and turbines to be re-processed and cleans the inner parts from me‐
chanical abrasion.
2.3 Safety instructions
2.3.1 General information
WARNING
Electrical power.
Electrical shock.
Damage to or malfunction of the unit.
▶ Connect the device to a properly grounded and fuse-protected power source only.
▶ Check if the information on the rating plate of the unit (voltage, frequency) is con‐
sistent with the available electrical power supply. The rating plate is attached on
the back panel of the device.
▶ Supply the device only with the voltage specified in the instructions for use, on the
rating plate, and on the device.
▶ Only use extension cords that are permissible for the indicated rated voltage and
are fitted with a protective conductor.
▶ Disconnect the unit from the power supply when you leave your practice.
11 / 36
Содержание QUATTROcare PLUS 2124 A
Страница 1: ...Instructions for use QUATTROcare PLUS 2124 A 1 008 3805 Always be on the safe side...