Instructions for use QUATTROcare PLUS 2124 A - 1.008.3805
5 Operation | 5.7 Chuck Servicing
28 / 36
A holder for accommodating the chuck service adaptor is provided in the side door.
The service adaptor is included in the accessories package at the time of delivery.
▶ Take the chuck service coupling out and attach it to coupling service point 4 (on
the far right). A MULTIflex adaptor must be mounted on the coupling.
▶ Press the handpiece, with the chuck to be serviced leading, against the tip of the
chuck service coupling.
Note
If this is done with insufficient pressure, there is a risk of the chuck service adaptor
being pushed off the service coupling during the servicing and of the servicing
agent leaking.
▶ Press the chuck service button ②.
ð
The spray is then briefly injected through the chuck, completing the service.
Quit the chuck service mode:
▪ Place the handpieces in the QUATTROcare PLUS 2124 A, close the front door
and start theservice procedure.
▪ Alternatively, after three minutes with no service procedure running, the device au‐
tomatically switches back to normal service mode.