▶ Then fill-up the water bottle with tap water.
Note
Substances remaining in the dosing head are not contamination, but rather
agents that are deposited there during extended periods of non-use after con‐
densation of the water. These deposits redissolve after several dosing process‐
es.
▶ Screw the water bottle tightly to the treatment centre in a clockwise rotation.
Apply compressed air to the water bottle.
3.2
Check the OXYGENAL 6 concentration
▶ Fill the tumbler with water.
▶ Use Merckoquant peroxide test strips 100 to 1,000 mg/l H
2
O
2
(Mat. no. 0.229.5003) to determine the OXYGENAL 6 concentration in the wa‐
ter germ reduction system. The OXYGENAL 6 concentration should be in the
range from 150 to 250 mg/l H
2
O
2
in normal operation.
Care instructions KaVo Primus® 1058 S/TM/C
3 Replenish and check the disinfectant | 3.2 Check the OXYGENAL 6 concentration
35/35