▶ Hold the handpiece at the taper sleeve and take off the cannula with a slight
twisting motion.
▶ Pull the gripping sleeve off the valve body.
Sterilization in a steam sterilizer in compliance with DIN EN 13060
▶ Weld the cannula and grip sleeve in sterile goods packaging.
▶ Sterilise with a triple fractionated vacuum (B class steriliser); hold for 4 mi‐
nutes; 134
°
C ± 1 °C; overpressure: 2.13 bar.
CAUTION
Damage caused by over-heating.
Damage to the gripping sleeve.
▶ Immediately remove the parts from the steriliser after the sterilisation cycle is
completed.
Storage
▶ Reprocessed products should be stored protected from dust with minimum ex‐
posure to germs in a dry, dark and cool place.
2.4.5
Preparation of the instrument hoses
Cleaning and disinfection of instrument hoses
▶ Clean the hoses and couplings with a cloth and water.
CAUTION
Damage caused by spray disinfection.
▶ Do not subject instrument hoses to spray disinfection.
▶ Subject instrument hoses to wipe disinfection.
Care instructions KaVo Primus® 1058 S/TM/C
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
14/35