Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550
4 Startup and shut down
14 / 28
4 Startup and shut down
WARNING
Hazard from contaminated products.
Infection hazard to the dentist and patient.
4
Prior to initial startup and after each use, reprocess the product and ac-
cessories.
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
4
Reprocess the product and accessories before disposal.
Also refer to:
7 Reprocessing steps in accordance with ISO 17664, Page 19
CAUTION
Damage from soiled and moist cooling air/compressed air.
Contaminated and moist cooling air can cause malfunctions.
4
Make sure that the supply of cooling air is dry, clean and uncontaminated
according to DIN EN ISO 7494-2.
Содержание MULTIflex Coupling 465 RN
Страница 1: ...Instructions for use MULTIflex Coupling 465 RN 0 553 1600 MULTIflex LUX Coupling 465 LRN 0 553 1550 ...
Страница 26: ......
Страница 27: ......
Страница 28: ...1 008 5363 bd 20211208 02 en ...