Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing
21 / 28
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess it in a washer disinfector only.
CAUTION
Automated reprocessing of couplings with high-pressure lamp.
Malfunction due to automated reprocessing, couplings with high-pressure lamp
are designed only for up to a max. of 50 reprocessing cycles.
4
Replace high-pressure lamp with KaVo MULTI LED lamp.
NOTE
Adapters are needed for automated cleaning.
Order adapter separately.
Also refer to:
8 Optional aids and consumables, Page 24
7.4.1 Preparation for automated internal and external
cleaning as well as internal and external disinfection
Miele Series G 7881/7891
Reprocessing with Miele AUF1/AUF2
adapter and silicone adapter ADS2
with a diameter of 16 mm
ADS2
AUF1/AUF2
7.4.2 Automated internal and external cleaning and
internal and external disinfection
KaVo recommends washer disinfectors in accordance with EN ISO 15883-1,
which are operated using alkaline cleaning agents.
The validation was performed in a Miele washer disinfector using the
"VARIO-TD" programme and the "neodisher MediClean forte" cleaner from Dr.
Weigert.
In addition, KaVo recommends the use of a neutraliser and a rinsing agent.
4
For programme settings and the adaptation options to be used, please refer
to the Instructions for Use of the washer disinfector.
7.4.3 Automated drying
The drying procedure is usually part of the cleaning programme of the washer
disinfector.