![KaVo KaVoLUX 540 LED U Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/kavo/kavolux-540-led-u/kavolux-540-led-u_technicians-instructions_3908096013.webp)
Technician's instructions KaVoLUX 540 LED U / T
2 Safety | 2.3 Safety instructions
CAUTION
Electrostatic discharge.
Destroys electronic components.
▶ Before you touch the electronic circuit boards, always earth them by touching the
earthing conductor (PE).
▶ When working on an open unit, takes precautions against ESD.
▶ Always hold electronic boards on their edge only.
Note
Switch off the light during treatment intervals to increase the service life of the LED
light.
Compliance with the national statutory regulations is obligatory during the use of the
device, in particular in as far as it concerns:
▪ the applicable health and safety regulations
▪ the applicable accident prevention regulations
The user is responsible for:
▪ exclusive use of equipment that is operating correctly
▪ protecting him- or herself, the patient and third parties from danger
▪ preventing any contamination by means of the device
The following regulations must be observed for the ceiling assembly of the KaVoLUX
540 LED T:
▪ Guidelines of the regulatory and building authorities
▪ Only use fastening elements (heavy-duty dowels) that are approved by the build-
ing authorities
▪ The electrical installation may only be performed by a certified electrician
▪ The electrical installation must comply with the requirements and guidelines set
down in VDE 0100-710
The following regulations must be observed when the KaVoLUX 540 LED T is moun-
ted to a treatment center:
▪ The electrical installation must comply with the requirements and guidelines set
down in VDE 0100-710
The safety checks (SC) must be done at 24-month intervals.
See also:
2
11 Safety checks - Testing instructions, Page 42