Technician's instructions KaVoLUX 540 LED U / T
11 Safety checks - Test instructions | 11.2 Ceiling mounting
48 / 56
Connecting the safety tester to the treatment centre without a KaVo
measuring cable
▶ Switch L + N of the on-site power supply cord to be voltage-free.
▶ Disconnect L + N on terminals X1.1 and X1.2.
▶ Connect the safety tester directly to terminals X1.1 (L) and X1.2 (N) and protective
earth conductor terminal (PE).
Note
The main switch of the ME device / ME system must be turned on during measure-
ment.
Connecting accessible conductive parts [ACP] to PE
ACP = accessible conductive parts
ACP
ACP
Note
Additional measuring points ACP X must be taken into consideration in the pres-
ence of accessories.
ACPs on the treatment light KaVoLUX 540 LED
No ACPs need to be connected to the operating light during the measurement with the
protective conductor (PE), as all relevant parts are connected to the protective con-
ductor (PE) before they leave the factory and included in the test.
Measuring the protective conductor resistance
Threshold:
< 0,3 Ω (maximum value!)