Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
▶
Clean the surface with a soft cloth and mild liquid cleaning agent.
Disinfection
Exterior disinfection by hand
WARNING
Highly flammable gases-.
Explosion.
▶
Allow product to cool before disinfecting.
NOTICE
Damage to the camera due to improper disinfection.
Irreversible destruction of the camera.
▶
Disinfect with wipes only, do not use disinfectant sprays!
▶
Do not immerse the camera in liquids!
▶
Do not sterilise the camera in an autoclave!
▶
Do not store the camera in Steri pack!
▶
Do not sterilise the camera with hot air or in a washer disinfector!
▶
Switch main device switch off.
▶
Allow the camera to cool down.
▶
Remove the camera.
▶
Disinfect the surface using a soft wipe and disinfectant.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
19 / 44