Care instructions Primus 1058 Life
3 Replenish and check the disinfectant | 3.1 Replenish OXYGENAL 6
38 / 44
Note
This allows the over-pressure to be released slowly from the water bottle.
Note
The water in the water bottle needs to be changed at least one daily. In order
to ensure the efficiency of the measures, KaVo urgently recommends repro-
cessing the water with KaVo OXYGENAL 6.
▶
For cleaning, rinse the water bottle with water at less than 50
o
C.
Filling the water bottle with the dosing facility.
▶
Wipe up any spilled OXYGENAL immediately in order to prevent discoloura-
tion of the surfaces.
▶
Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing
facility.
▶
Guide the tube through the orifice of the coarse sieve of the KaVo OXY-
GENAL 6 bottle.
▶
Screw the KaVo OXYGENAL 6 dosing facility tightly to the KaVo OXYGENAL
6 bottle.
▶
Rotate the orifice in the lid to a filling level of 1.5 litres.
ð
Lid snaps-in in the filling position.
▶
Place the KaVo OXYGENAL 6 bottle on a level surface.
▶
Squeeze the KaVo OXYGENAL 6 bottle until the germ reduction liquid rises
above the red marker.
ð
In response to over-filling, the excess quantity above the tick mark is aspir-
ated automatically to ensure exact dosing.
▶
Place the water bottle in the KaVo OXYGENAL 6 bottle. Make sure that the
spout of the KaVo OXYGENAL 6 dosing facility (nose) is situated right above
the opening of the water bottle.