Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
▶
Pull the gripping sleeve off the valve body.
Sterilization in a steam sterilizer in compliance with DIN
EN 13060
▶
Weld the cannula and grip sleeve in sterile goods packaging.
▶
Sterilise with a triple fractionated vacuum (B class steriliser); hold for 4
minutes; 134
°
C ± 1 °C; overpressure: 2.13 bar.
NOTICE
Damage caused by over-heating.
Damage to the gripping sleeve.
▶
Immediately remove the parts from the steriliser after the sterilisation cycle
is completed.
Storage
▶
Reprocessed products should be stored protected from dust with minimum
exposure to germs in a dry, dark and cool place.
2.4.5 Preparation of the instrument hoses
Cleaning and disinfection of instrument hoses
▶
Clean the hoses and couplings with a cloth and water.
NOTICE
Damage caused by spray disinfection.
▶
Do not subject instrument hoses to spray disinfection.
▶
Subject instrument hoses to wipe disinfection.