Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
14 / 44
▶
Take the roller out of the instrument arm and then place the hose in the
germ reduction attachment.
▶
Pull-off the swing-arm in an upward direction.
▶
Clean the swing-arm and roller under running water.
▶
Wipe the instrument arm S table with disinfectant.
Sterilisation of instrument arm S table
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise the instrument arm S table.
2.4.4 Setting up the 3-function and multifunction
handpiece
Note
Wet the O-rings of the gripping sleeves after sterilisation using silicone
grease (
Mat. no. 1.000.6403
) or KaVo Rota Spray 2 (
Mat. no.
0.411.7510
). After replacing the cannula, the air channel must be blown out
before the start of treatment to ensure that the air exiting during treatment
is dry.
Cleaning the triple-function and multifunctional
handpiece
Automatic cleaning
Not applicable.
NOTICE
Damage caused by improper cleaning/disinfection.
▶
Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶
Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶
Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.