Short instructions for use 10 Hygiene
28
S-table
▶ Lock the swing-arm into the front catching position to ensure that the handpiece is removed.
▶ In the case of swing-arms without a catching function, insert the swing-arm lock (Mat. no.
1.008.1511) to ensure that the handpiece is removed.
▶ Hold the 3-way or multifunctional handpiece at the gripping sleeve and take off the cannula with a
slight twisting motion.
▶ Insert the triple-function or multifunctional handpieces without the cannula into the clamping de‐
vice of the germ reduction attachment. Make sure that the "W" button (water) of the triple-function
or multifunctional handpiece is activated, and that no air is exiting.
▶ Place the instrument hoses of the dentist and assistant elements in the corresponding places of
the germ reduction attachment.
▶ Support tubing with the enclosed holder.
Note
All water regulation mechanisms for the motor, instruments, and multiflex couplings
must be open.