11.2 Additional warning messages
Malfunction
Cause
Remedy
A beep is issued every
ten seconds and the "In‐
tensive germ reduction"
LED (green) flashes.
The Oxygenal container
is empty.
▶ Refill the Oxygenal container.
Ten beeps are issued.
The Oxygenal container
is too full.
▶ Stop filling the Oxygenal container.
A melody sounds.
The amalgam separator
CAS1 is almost full
(95%).
▶ Exchange the amalgam container.
The CAS1 amalgam
separator is defective.
▶ Also refer to: Instructions for use for the CAS
1 or
▶ Call a Service technician.
LED on "HYDROclean"
button (red) flashes.
Malfunction in the amal‐
gam separator.
▶ Call a service technician.
▶ Note the amalgam separator warning notice.
Also refer to: Operating instructions of the
amalgam separator
See also:
2
Operating instructions of the amalgam sepa‐
Emergency shut off of
the bowl valve (only
when external suction is
installed)
▶ Call a technician.