Short instructions for use 12 Service and error messages in the stand-by menu
35
12 Service and error messages in the stand-by menu
If a status message is available, the standby menu shows an exclamation mark on selection key "S2"
①.
1
▶ Press the "S2" selection key ① to display status messages.
▶ Press the selection keys "+" ② and "-" ① to switch between multiple status messages.
▶ Press the "OK" selection key ③ to exit from the display of status messages.
Error messages in the status display
Malfunction
Cause
Remedy
Display shows: ID 33
No CAN node and/or in‐
ternal communication is
faulty.
▶ Call a service technician.
Display shows: ID 64
Water is shut off.
▶ Turn water on.
Water works leaks
strongly.
Water works malfunc‐
tion
▶ Call a technician.
Display shows: ID 65
Safety switch of bowl
suction has been
reached.
▶ Turn external suction on.
▶ Check and clean, if required, the bowl valve.
Display shows: ID 66
Amalgam separator
malfunction
▶ Remedy the malfunction.
See also:
2
Instructions for Use of amalgam separator
Display shows: ID 67
The oxygenal container
is empty.
▶ Refill the Oxygenal container.
See also:
2
Display shows: ID 68
Call for service
▶ Have a service performed.
▶ Call a technician.