▶ Remove the chuck ①.
▶ Insert a new tool or test probe ① into the chuck as far as it will go.
▶ Grip the handpiece part 2, and turn the motor 3 in the direction indicated by
the arrow until the chuck closes completely.
▶ Slowly turn the shaft in the pin ① until the clamp B can be inserted with a
slight amount of resistance into the grip tip ②.
▶ Open the chuck.
Shove out the chuck with the original chuck key A while the pin or tool is inser‐
ted.
▶ Clean the front opening of the handpiece with a brushMat. no. 0.411.3180.
▶ Clean the chuck thread and surface as well as the rotor cone, and apply a
slight amount of lubricant.
Instructions for use K-POWERgrip Handstück 4941
6 Service and Care | 6.1 Cleaning
18/29