Technician's instructions GENTLEpower LUX Contra-angle 25 LP - REF 1.001.2849
4 Testing | 4.1 Testing the GENTLEpower LUX Angle Piece 25 LP
18 / 20
Testing
Test equipment
Procedure
Check concentricity
▪ Test drill
Run the angle piece with test drill
off-load at 200,000 rpm.
There should be no noticeable vi‐
brations or running noise.
Check power consumption
▪ Test drill
Measure the power consumption
under test conditions with head and
test drill.
Power consumption: max. 150 mA
Check seals
▪ Leakage air hose
Pressurize the media chamber with
compressed air
(2 bar) and pressure-test the spray
drilled hole for 15 seconds.
The compressed air must remain
constant.
Run a spray test
▪ Spray test drill
Carry out a visual inspection of the
spray mist.
The spray jets should hit the ball of
the spray test drill.
▪ Test device
Measure air and water quantity un‐
der test conditions.
Air quantity: at least 2.5 Nl/min.
Water quantity: at least 50 ml/min.