▶ Place the contra-angle handpiece in a hot-air furnace and allow the adhesive 30
minutes
to set at 120
o
C.
▶ Remove the head casing ⑤ from the sleeve ⑧ again.
▶ Use Loctite 306 adhesive to fix the inside of the spray sleeve ⑧ to position ⑥ and
the fibre-optic bundle ④ making sure that the fibre-optic bundle is situated such as
to be flush at the front. Remove any residual adhesive from the end of the fibre-
optic bundle.
▶ Place the angle piece in a hot-air furnace and allow the adhesive to set for 30 mi‐
nutes at
120
o
C.
3.1.6 Assembling the head
Tool
Description
Wrench (Mat. no. 1.001.3288)
Universal bit holder (Mat. no. 1.002.4577
Torque handle (Mat. no. 1.002.4578) with
100 Ncm
Circlip pliers (Mat. no. 0.411.2421)
▶ Fit one O-ring ② in the groove on the spray insert ③ and one in the attachment of
the spray insert ③.
▶ Insert the spring washer ④ and drive axle ⑤ into the spray insert ③.
▶ Insert the spray insert ③ into the head casing ①.
▶ Screw the cover ⑥ into the head casing ① and tighten with a torque of 100 Ncm
with the wrench, the universal bit holder and the torque handle.