Instructions for use EXPERTmatic E10 C – 1.007.5560
2 Safety | 2.6 Service and repair
9 / 28
4
Have the service and testing tasks carried out according to the Medical
Device Operator Ordinance.
4
Following expiry of the warranty, have the tool holding system checked
once a year.
4
KaVo recommends specifying in-house service intervals where the medical
device is brought to a professional shop for cleaning, servicing and a func-
tion check. Define the service interval depending on the frequency of use.
As a result of the use of NON-KaVo original spare parts during the repair, parts
may become detached and injure patients, users or third parties. This may res-
ult in aspiration, swallowing of parts and possibly even a risk of suffocation.
4
Only use spare parts that comply with the specification for repair; original
manufacturer spare parts comply with the specification.
NOTE
If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.
The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by MPG (in English medical device law) §4, section 1 No. 1 and there-
fore requires its own conformity check.