Instructions for use EXPERTmatic E10 C – 1.007.5560
6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions
17 / 28
6 Checking for malfunctions and troubleshooting
6.1 Checking for malfunctions
CAUTION
Overheating of the device.
Burns or product damage from overheating.
4
If the device overheats, stop working and have the service personnel repair
the device.
CAUTION
Missing or damaged O-rings.
Malfunction and premature failure.
4
Make sure that all O-rings are on the coupling and are undamaged.
4
If the medical device overheats while idling, check the cooling air flow.
4
If the medical device overheats when exposed to load, the medical device
needs to be serviced.
4
When the speed drops or is uneven, the medical device needs to be ser-
viced.
4
If there is no O-ring on the motor coupling, replace the O-ring.
Also refer to:
Instructions for use of motor
6.2 Troubleshooting
WARNING
Repair WITHOUT using KaVo original spare parts.
Parts could come loose and cause injury. Aspiration, swallowing of parts and
danger of suffocation.
4
Only use spare parts that comply with the specification for repair; original
KaVo spare parts comply with the specification.
NOTE
If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.
The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by MPG (in English medical device law) §4, section 1 No. 1 and there-
fore requires its own conformity check.
6.2.1 Replacing the O-rings on the motor coupling
CAUTION
Improper care of the O-rings.
Malfunction or complete failure.
4
Do not use Vaseline or other grease or oil.
4
Spray a lint-free cloth with KaVo Spray and apply it to the O-rings on the
coupling.
4
Press the O-ring between your fingers to form a loop.
4
Push the O-ring to the front, and remove it.
Содержание EXPERTmatic E10 C
Страница 1: ...Instructions for use EXPERTmatic E10 C 1 007 5560...
Страница 28: ...1 008 1787 bd 20240411 03 en...