Instructions for use EXPERTmatic E10 C – 1.007.5560
7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.3 Automated reprocessing
21 / 28
Cleaning with KaVo CLEANspray (internal and external)
Carry out non-protein-fixing internal cleaning with a CLEANspray and external
cleaning with a cloth soaked with CLEANspray:
4
Attach fitting adapter to the CLEANspray can.
4
Attach instrument with the drive side to the adapter.
4
Rinse through the instrument for at least 3 x 2 seconds (solution must flow
out visibly).
4
Take the instrument from the adapter and set it down.
4
Spray the surface of the instrument with CLEANspray until the entire sur-
face is wet.
4
Allow the cleaning agent to act for 1 minute.
Manual disinfection
Disinfection with WL-cid from Alpro (in external)
Carry out internal disinfection with WL-cid and external disinfection using a
cloth soaked in WL-cid:
4
Attach fitting adapter to the WL-cid can.
4
Attach instrument with the drive side to the adapter.
4
Rinse through the instrument for at least 2 x 3 seconds (solution must flow
out visibly).
4
Take the instrument from the adapter and set it down.
4
Spray the surface of the instrument with WL-cid until the entire surface is
wet.
4
Allow the disinfectant to act for 5 minutes.
7.2.3 Manual drying
Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts.
4
Cover the medical device with the KaVo Cleanpac bag, and place it on the
corresponding care adapter.
4
Hold the can vertically.
4
Press the spray key for at least 3 seconds.
Also refer to:
Instructions for use KaVo DRYspray
4
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
Also refer to:
7.4 Care products and systems - Servicing, Page 22
7.3 Automated reprocessing
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.