Instructions for use EXPERTmatic E10 C – 1.007.5560
5 Operation | 5.3 Inserting the dental bur
15 / 28
5.3 Inserting the dental bur
NOTE
Only use carbide burs or diamond burs that conform to DIN EN ISO 1797 type
1 and type 2, are made of steel or hard metal, and meet the following cri-
teria:
- Shaft diameter: 2.334 to 2.350 mm
with a drill bit stop:
- Shaft clamping length: at least 12 mm
- Overall length: max. 22 mm
without a drill bit stop:
- Shaft clamping length: at least 30 mm
- Overall length: max. 44.5 mm
WARNING
Use of unauthorised dental burs.
Injury to the patient or damage to the medical device.
4
Comply with the instructions for use and the intended use of the dental bur.
4
Only use dental burs that do not deviate from the specified data.
CAUTION
Do not use damaged dental burs.
Risk of injury from swallowing a dental bur that falls out.
4
Do not use damaged dental burs.
4
Do not use dental burs that have been hit.
4
Do not use dental burs with adherent soiling.
4
Do not use visibly imbalanced dental burs.
CAUTION
Contaminated, sharp-edged dental bur.
Patients, users or third parties could be infected by contaminated medical
devices.
4
Use gloves or a finger guard when you test, insert and remove the dental
bur.
CAUTION
Dental bur with damaged, worn or deformed shafts.
A damaged product or damaged or NOT KaVo original components can injure
patients, users or third parties.
Injury hazard, dental bur may fall out during treatment.
4
Never use dental burs with damaged, worn or deformed shafts.
CAUTION
Defective chuck system.
Injury hazard, dental bur may fall out during treatment.
4
Pull on the dental bur to check if the chuck system works properly and if the
dental bur is firmly clamped.