Assembly instructions Everest
®
engine Base Camp 4140
7 Calibrate the Everest engine Base Camp 4140
|
7.3 Evaluate the calibration body
Index Value
Description
Required action
1
-32100
Value for axis X
Do not change this va‐
lue.
2
16077
Value for axis Y
Enter in the form under
Y
0,Alt
.
3
-454
Value for axis Z
Do not change this va‐
lue.
4
-189970
Value for axis A
Do not change this va‐
lue.
5
-115754
Value for axis B
Do not change this va‐
lue.
▶ Enter the value of index 2 in the form under Y
0,OLD
.
▶ Close the window by clicking the button "End" (F8).
Record P51
OLD
:
▶ Switch to the "DATA" menu.
▶ Click the button "Data selection" (F1).
▶ Click the "Parameter" button (F2) and then the "DATA" menu.
▶ Click the button "Change" (F5).
43/51