Care instructions ESTETICA E50 Life
3 Replacing and checking cleaning and germ reduction agents | 3.3 Replacing the DEKASEPTOL Gel bottle
(DEKAmat assembly kit only)
54 / 58
H2O2 concentration
Measuring cup
Ratio water : treated
water
Value shown on test
strip
0.02 % (200 ppm)
standard concentration
40 ml
39:1
5 mg/l
0.05 % (500 ppm)
100 ml
99:1
5 mg/l
0.1 % (1,000 ppm)
40 ml
39:1
25 mg/l
0.2 % (2,000 ppm)
80 ml
79:1
25 mg/l
0.25 % (2,500 ppm) in-
tensive germ reduction
concentration
100 ml
99:1
25 mg/l
0.3 % (3,000 ppm)
120 ml
119:1
25 mg/l
3.3 Replacing the DEKASEPTOL Gel bottle (DEKAmat
assembly kit only)
Note
The DEKASEPTOL gel bottle can be replaced only in combination with the
DEKAmat assembly kit.
Requirement
The device must be turned on when changing DEKASEPTOL Gel bottle.
▶
Turn the device on.
▶
Open the door in the base of the unit and take out the DEKASEPTOL Gel
bottle.
▶
Unscrew the lid of the new DEKASEPTOL Gel bottle.
▶
Unscrew the connector of the empty DEKASEPTOL Gel, screw the connector
to the new DEKASEPTOL Gel bottle and insert the bottle.
Accessories: DEKASEPTOL refill canister (
Mat. no. 1.000.7205
) with drain
tap (
Mat. no. 0.224.6876
)