Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
Cleaning of handpieces
Manual external cleaning
▶
Unscrew sealing cap and fibre optic conductor.
▶
Gently clean the individual parts under running water using a soft brush or
a soft cloth.
▶
Rinse the external housing of the handpiece thoroughly.
Manual internal cleaning
▪ Distilled, deionised water
(aqua purificata as specified in Pharm. Eur. or USP)
- with microbial count < 10 cfu/ml or sterilised
- with sufficiently low endotoxin and particle concentration
▶
Attach disposable syringe to the back nozzle.
▶
Rinse in the normal flow direction, do not rinse against the flow direction.
▶
If an aldehyde-free cleaning and disinfection solution is used, subsequently
rinse at least thrice with distilled or deionised water.
Automated external cleaning
KaVo recommends washer disinfectors from Miele in accordance with EN ISO
15883-1 that are operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD programme, the cleaning
agent neodisher® MediClean and the neutralisation agent neodisher® Z.
▶
For programme settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the washer disinfector.
▶
In order to prevent negative effects on the KaVo medical device, make sure
that the interior and the exterior of the medical device are dry after comple-
tion of the cycle, and then lubricate it immediately with care products from
the KaVo care system.
Automated internal cleaning
Possible only with devices providing the option or function of intensive germ re-
duction.
▶
Leave the handpiece on the device and carry out the intensive germ reduc-
tion (see Instructions for Use of the respective treatment unit).