Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system
34 / 58
CAUTION
Health damage due to germ formation.
Infection hazard.
▶
Before starting, rinse all the water drain lines without instruments.
▶
Before start-up and after the device has not been used for a while (week-
ends, holidays, vacations, etc.), rinse or purge with air the air and water
lines.
▶
Carry out an intensive germ reduction.
▶
Actuate the tumbler filler repeatedly.
The water-conducting system includes all water ducts of the treatment centre
to which the patient may be exposed. This includes the instruments on the
dentist unit and assistant unit as well as the tumbler filler on the spittoon bowl
of the treatment centre.
During normal operation of the treatment centre, the continuous germ reduc-
tion function secures continuous germ reduction in the water conducting sys-
tems due to the automatic dosing of the germ reduction liquid, KaVo OXY-
GENAL 6.
Rinsing programme and intensive germ reduction facilitate the rinsing and germ
reduction of the water-supplying system. This counteracts the formation of mi-
cro-organisms at times of stagnation and ensures a hygienic water supply for
the patients.
2.8.1 Rinsing programme
The rinsing programme 02 provides for automatic rinsing of the water-conduct-
ing system and all the instruments including the tumbler filler for 2 minutes
each (hydrogen peroxide concentration 0.02 %). In the short rinsing pro-
gramme 01, all instruments on the dentist unit and assistant unit are rinsed for
20 seconds each.
Procedure of the rinsing programme:
▪ Before you start working (rinsing programme 02)
▪ After each patient and in the evening (rinsing programme 01)
▶
Briefly press the "Intensive germ reduction" button.
Preparation for rinsing programme and intensive germ
reduction
▶
Position the germ reduction attachment on the porcelain spittoon bowl.
with patient part assembly kit / without patient part assembly kit