Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys-
tem
42 / 58
ter (limited rinsing). The water is aspirated from the Aquamat through the filter
housing to the separation system. The spittoon bowl rinsing is alternately
turned on and off.
The HYDROclean function prevents the coagulation and adhesion of proteins in-
side the drainage system and improves the transporting of fine particles.
Carrying out the HYDROclean function:
▪ After each treatment
Preparation
Requirement
Suction hoses remain hanging in the holder.
▶
Press the "HYDROclean" button.
ð
The HYDROclean function is in the preparatory stage.
Start the HYDROclean function
▶
Press the "HYDROclean" button.
ð
This starts-up the HYDROclean function.
The HYDROclean function is completed when the top LED goes dark.
The process can be discontinued at any time by pressing the "HYDROclean"
button.
2.9.3 Manual internal cleaning of the suction hoses at
the extraction point (optional)
Note
Manual internal cleaning of the suction hoses can be done only in combina-
tion with the CENTRAmat / DEKAmat assembly kit.
For manual cleaning, a suction hose is held to the extraction point. The suction
hose, drain and separation system are rinsed. The water is sucked out of the
Aquamat through the extraction point and suction handpiece to the separation
system.
Carrying out the manual inner cleaning
▪ After each treatment
▶
Touch all suction hoses consecutively against the extraction point such that
no ambient air is aspirated.