2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys-
tem
2.9 Reprocessing and automatic hygiene functions of
the suction system
2.9.1 Reprocessing of the suction system including
suction hoses
Cleaning and disinfection of the suction system
including suction hoses
Note
Only DEKASEPTOL Gel may be used for cleaning and germ reduction of the
suction system
Note
Clean the suction hoses after each treatment and disinfect them with DEKA-
SEPTOL Gel daily.
Application of DEKASEPTOL Gel
▶
Aspirate one tumbler of cold water with each suction hose.
▶
With the suction handpiece attached or by hand, press down the DEKA-
SEPTOL dosing facility once each for the saliva ejector hose and the spray
mist ejector.
▶
Aspirate the DEKASEPTOL Gel from the dosing facility.
▶
Aspirate another tumbler of cold water with each suction hose.
▶
Finally, aspirate another dose of DEKASEPTOL Gel and allow it to act.
▶
Place the suction hoses in the holder.
DEKASEPTOL Gel basic set
DEKASEPTOL Gel refill canister
Mat. no. 10007204
Mat. no. 10007205
2.9.2 HYDROclean function
The HYDROclean function is an automatic programme for hydro-mechanical
cleaning of the drainage and separation systems within and outside the treat-
ment unit. The cleaning process is carried out by the supply of fresh, clear wa-
41 / 58