Note
Do not pull the dentist unit by the instrument hose.

▶ To adjust the dentist unit height, release the brake, adjust the height, and re‐

set the brake.

5.14

   

Changing the height of the assistant element (optional assem‐

bly kit)

Note
The "height-adjustable assistant element" is an optional assembly kit. If the op‐

tional assembly kit has been selected, the assistant element can be positioned at

three vertical levels.

▶ To set a higher level, pull the assistant unit upward gently until it audibly locks

in place.

▶ To set a lower level, pull the assistant unit all the way up until the lock relea‐

ses, and then lower the assistant unit.

5.15

   

Position the patient simulator (pneumatic adjustment)

 

 

 

 

 

 

 

Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH
5 Operation | 5.14 Changing the height of the assistant element (optional assembly kit)

34/48

Содержание DSEclinical 5197

Страница 1: ...Instructions for use DSEclinical Typ 5197 DSEclinical Typ 5197eH Always be on the safe side...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...lable with TM or S table 11 3 2 Identification plates 13 3 3 Technical Specifications 14 3 4 Scope of delivery 15 4 First use 17 4 1 Fitting and connecting 17 5 Operation 20 5 1 Operation in general 2...

Страница 4: ...it 34 5 15 Position the patient simulator pneumatic adjustment 34 5 16 Electrically position the patient simulator optional kit 36 5 16 1 Adjusting the patient simulator gradually 36 5 16 2 Save the p...

Страница 5: ...on required 1 2 Service KaVo Customer Service 49 0 7351 56 2500 Service Einrichtungen kavo com Please refer to the serial number of the product in all inquiries For further information please visit ww...

Страница 6: ...aging regulations employing waste management or recycling compa nies Comply with the comprehensive return system KaVo has had its sales packaging licensed for this purpose Please comply with the regio...

Страница 7: ...t use the product If the product is damaged but there was no discernable damage to the packaging upon delivery proceed as follows 1 Report any damage to the shipping company either immediately or no l...

Страница 8: ...Permissible stacking load Temperature range Humidity Air pressure Instructions for use DSEclinical Typ 5197 DSEclinical Typ 5197eH 1 User instructions 1 4 Transportation and storage 6 48...

Страница 9: ...cument to prevent personal and property damage CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moderate injuries WARNING WARNING indicates a hazardous situ...

Страница 10: ...make a decision whether nor not further individual tests are required in order to attain the safety target If safety deficiencies or damage is are detected during the re test the testing process must...

Страница 11: ...n hazard DANGER Electricity Electrical shock Before opening the unit covers pull the plug from the mains or disconnect the unit from the voltage source WARNING Injury or damage from damaged functional...

Страница 12: ...uenced by electromagnetic fields Clarify with the user before starting work 2 3 2 Product specific CAUTION Danger if unit not in proper working order or not being used properly Risk of injury and dang...

Страница 13: ...tist element TM DentSim optional accessories Dentist element TM Assistant s element Main fuse Patient Simulator Main switch Foot control Instructions for use DSEclinical Typ 5197 DSEclinical Typ 5197e...

Страница 14: ...e for both units with the exception of the electrical settings for the type 5197 eH Therefore when describing the same functions and settings only type 5197 will be mentioned Instructions for use DSEc...

Страница 15: ...nical 5197 eH 120 V Rating plate for the DSEclinical 5197 230 V and rating plate for the DSEclinical 5197 eH 230 V Type Type of device DSEclinical type 5197 SN Serial number REF Material number CE mar...

Страница 16: ...0 Hz Mains fuse Type T6 3 H 250 V Maximum permissible mains voltage fluctuations 10 of the nominal voltage Overvoltage category II Media Water quality Tap water Water hardness 1 5 to 2 14 mmol l 8 4 t...

Страница 17: ...ove mean sea level Transportation and storage conditions temperature 20 to 55 C Relative humidity 5 to 95 non condensing Air pressure 700 hPa to 1060 hPa 3 4 Scope of delivery DSEclinical type 5197 ba...

Страница 18: ...3825 Spray mist extractor Description Material number Secretion hose Mat no 0 211 1761 Suction canula Mat no 0 236 2127 Saliva ejector Description Material number Saliva ejector hose Mat no 0 211 1571...

Страница 19: ...guaranteed under all circumstances Comply with the relevant regulations when laying mains circuits and IT cir cuits CAUTION Cable or plug connections can be damaged when connecting the devices Use in...

Страница 20: ...See also Instructions for Use of instrument or motor Fitting triple and multifunction handpiece CAUTION Damage due to missing media Air and water heating systems are destroyed Check if the air and wa...

Страница 21: ...through Connect axially handpiece hose and handpiece Insert the rubber dampers into the gripping sleeve of the handpiece Screw hose casing to valve body Instructions for use DSEclinical Typ 5197 DSEcl...

Страница 22: ...ient simulator into the treatment position See also 5 15 Position the patient simulator pneumatic adjustment Page 34 5 16 Electrically position the patient simulator optional kit Page 36 Remove the de...

Страница 23: ...treatment unit with a common media supply this treatment unit must first be turned on Only then is the device connected to the media 5 1 2 Switching off and tidying up the DSEclinical type 5197 5197 e...

Страница 24: ...the footswitch out of the range of movement 5 2 Overview of controls 3 7 2 1 5 4 6 8 9 10 11 Overview operating elements with dentist element TM Operating light Control panel Satelec Mini LED Camera T...

Страница 25: ...features of the dentist or assistant unit can differ from the picture The dentist element can be equipped with four or five nozzles optionally with tur bine air motor micromotor KL 701 KL 703 and PIEZ...

Страница 26: ...ovements The instrument hoses must not touch the floor in the parked position 5 3 Control panel Dentist element TM S Yellow LED shining device is ready for operation memory level dentist 2 is activate...

Страница 27: ...ng instructions for use service instructions and installation instructions in the instrument packaging The following settings can be changed Speed cooling condition cold light intensity Remove the tur...

Страница 28: ...zer sounds A buzzer is equal to one level 5 5 5 Storing values To store settings hold the LP AP button with the instrument removed until you hear a signal 5 6 Use INTRA LUX Motor K 200 and INTRA LUX M...

Страница 29: ...ltifunction foot control The LED for the motor s rotational direction shines Anticlockwise rotational direc tion is preselected 5 6 2 Storing values To store settings hold the LP AP button with the in...

Страница 30: ...8 months at the latest Using the torque wrench provided with the accessories screw the instru ment insert into the handpiece and tighten it until the maximum torque is reached This is signaled when th...

Страница 31: ...handpiece cool after long use Turn off unit Unscrew the hose ferrule from the handpiece and remove the handpiece from the hose coupling Pull the high pressure lamp out of the socket Insert the new hi...

Страница 32: ...e cannula is locked into place and firm ly seated Only use original KaVo cannulas Note Cannulas can be rotated by 360 On time is 5 minutes while the resting time is 3 minutes Remove the turbine from t...

Страница 33: ...ld light for the multifunction handpiece have been preselected To store settings hold the LP AP button with the instrument removed until you hear a signal 5 9 Using Satelec mini LED See also Satelec M...

Страница 34: ...for positioning chair back The instrument is started by pressing the footplate If the footplate is re leased the instrument stops By swivelling the footplate to the right or left the speed can be gra...

Страница 35: ...order to achieve the highest possible flexibility in the ergonomic positioning of the dentist element for right handed and left handed users Move and position the dentist element carefully Note Set t...

Страница 36: ...mbly kit has been selected the assistant element can be positioned at three vertical levels To set a higher level pull the assistant unit upward gently until it audibly locks in place To set a lower l...

Страница 37: ...amaged Press brake button The swivel arm mechanism lock is released Set the height and inclination of the patient simulator To do this carry the pa tient simulator to the chair back with both hands Le...

Страница 38: ...is activated the LED on the SP button on the dentist unit starts to flash CAUTION Adjustment of the patient simulator position is associated with a risk of getting pinched Various parts of the height...

Страница 39: ...an be saved and retrieved at any time by the press of a button Win the position is retrieved the patient simulator automatically moves to the saved position the so called automatic position or AP for...

Страница 40: ...ch Press the AP 0 AP 1 AP 2 or SP buttons on the control element until you hear a beep The selected automatic position is saved on the footswitch Retrieve the saved patient simulator positions CAUTION...

Страница 41: ...ntrol are blocked The blocking can be removed by briefly pressing the stirrup switch The functions are then available Spray preselection AP footswitch Foot pedal Footswitch blowing air AP optional The...

Страница 42: ...s into the automatic position 5 17 Adjust head rest Adjust bar length Push head rest in and out to change bar length Setting inclination Turn clamping button to the left Bring head rest into position...

Страница 43: ...he patient simulator The head of the dental patient simulator can perform the following movements Approximately 25 forward tilt Approximately 40 backward tilt Approx 20 inclination to left and right A...

Страница 44: ...e dental patient simulator Put the head in the required position and adjust the headrest accordingly After making the adjustments tighten the screw on the head again Note The brake for the neck can be...

Страница 45: ...ent The patient simulator is is in the lowest lying position Pivot the dentist and assistant unit above the patient simulator Instructions for use DSEclinical Typ 5197 DSEclinical Typ 5197eH 5 Operati...

Страница 46: ...fixation of camera Fixation spring arm DentSim Camera made by Image Navigation Swivel arm DentSim simulates the motion of handpieces on an external screen Loosen the screws and position the camera See...

Страница 47: ...adjusting motor once a year for proper and correct connection 6 2 Care If dirty clean the unit on the outside with a damp cloth Clean the cranium denture and face mask of the patient simulator See als...

Страница 48: ...h is off Turn on main switch No voltage supply Main fuse is blown Unplug the unit from the mains Change the microfuse T 6 3 H Mat no 0 223 2783 The main fuse is next to the main switch Cold light not...

Страница 49: ...ce all O rings in the multi flex coupling Suction hoses not working Slides on the bodies are closed Open the slide valve Sieves in suction connector blocked Exchange the sieve Foot switch for vacu sto...

Страница 50: ...ing on model Intraoral camera ERGOcam 4 depending on model Dental patient simulator cranium denture Figure Description Material number G 50 Mat no 0 622 0521 G 50 J Mat no 0 622 0581 G 50 JM Mat no 0...