4
First use
4.1
Fitting and connecting
Note
The DSEclinical type 5197 dental simulation unit may be installed only by trained
technical staff and must be bolted to the floor in accordance with KaVo stand‐
ards.
The technicians' and fitting instructions for the assembly, setting up and fitting
are to be complied with.
CAUTION
Damage caused by incorrectly laid mains circuit and IT circuit.
Incorrectly laid mains and IT circuits can render the use in accordance with the
intended purpose of the unit to be no longer guaranteed under all circumstances.
▶ Comply with the relevant regulations when laying mains circuits and IT cir‐
cuits.
CAUTION
Cable or plug connections can be damaged when connecting the devices.
Use in accordance with the intended purpose is no longer guaranteed under all
circumstances.
▶ Proceed with great caution and care when connecting devices.
Requirements of the mains connection cable:
▪ Withstand voltage: 10% above the rated voltage of the supply network
▪ Cross-section: compliance with the country-specific requirements is required,
but at least 1.5 mm
2
▪ Temperature range: at least 70 °C
▪ Conductor: L1, N, PE
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH
4 First use | 4.1 Fitting and connecting
17/48