① PE
③ N
② L1
Requirement
The air to be used for the dental simulation unit must be dry, oil-free, non-conta‐
minated, and free of pollution. Use a compressor with a dry air system if re‐
quired.
▶ Connecting the voltage supply and air connection.
▶ Depending on the model, connect the water connection and connection to the
central suction unit.
▶ Install motors and instruments as outlined in the instructions.
See also: Instructions for Use of instrument or motor.
Fitting triple and multifunction handpiece
CAUTION
Damage due to missing media.
Air and water heating systems are destroyed.
▶ Check if the air and water are connected.
▶ Check the air and water supply!
▶ If possible, switch the heating off at the unit when putting into operation for
the first time or after servicing! Press the buttons carefully several times until
the media are available. Then activate heating and check its operation.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH
4 First use | 4.1 Fitting and connecting
18/48