![KaVo ABSORmatic plus Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/kavo/absormatic-plus/absormatic-plus_operating-instructions-manual_3769077033.webp)
A 11 Operating faults
Repairs and servicing work on the
electrical part of this equipment must
only be undertaken by skilled personnel or
by persons trained in our factory who are
aware of the safety regulations. Disconnect
the power plug from the power supply or
open the circuit breaker respectively before
opening any part of the equipment.
Error messages are optically indicated on
the screen
.
11.1 Error message
Error 10 - (
1200
) Filter rotation
I
• No filter movement during cleaning
process.
C
• a) no filter inserted
b) cleaning motor faulty
c) filter position sensor faulty
R
• a) insert filter
b) and c) consult customer service dept.
Request service.
11.2 Error message
Error 11 - (
1200
) Dust container
I
• Dust container has been pulled out or
microswitch has malfunction.
C
• a) Dust-container has not been pulled
back after cleaning.
b) Micro switch defect or dirty.
R
• a) Correct re-location of dust container
b) clean or replace control cam of the
micro switch.
11.3 Error message
Error 23 - (
1200
) EEPROM defect
I
• Component EEPROM defect
C
• The EEPROM is tested if unit is swit-
ched on through power switch. This
error appears if the component is faulty
or indicates any malfunction at times.
After identification with pushbutton
"Enter"
, the process continous. In
case of malfunction, the specific custo-
mer datas cannot be read or stored.
R
• Press pushbutton
"program stored". If
fault is repeating - consult customer ser-
vice departement.
I
= Instructions
C
= Cause
R
= Remedy
31
ABSORmatic
®
plus.
ABSORmatic®
ABSORm
atic
®
ABSORmatic®
Û