![KaVo ABSORmatic plus Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/kavo/absormatic-plus/absormatic-plus_operating-instructions-manual_3769077030.webp)
28
ABSORmatic
®
plus.
A 10 Maintenance
Repairs and servicing work on the
electrical part of this equipment must
only be undertaken by skilled personnel or
by persons trained in our factory who are
aware of the safety regulations. Disconnect
the power plug from the power supply or
open the circuit breaker respectively before
opening any part of the equipment.
10.1 Cleaning
Rub or wipe away dirt from the outside of
the dust collector with a damp cloth.
When cleaning, protect the dust col-
lector against penetration by water.
The manufacturer or authorized person
must carry out an inspection of the dust
collection system, e.g. to check for damage
to the suction system, leak-tightness of the
equipment and function of the monitoring
devices.
10.2 LED display of the “Filter“ mode
operating
lights up (see operating
faults
11.7 "Warning 30 - Filter cleaning"
(1201)
.
The filter is dirty and will be cleaned of the
next operational interruption.
A short filter cleaning with filling level
defection fakes place automatically when
the power switch
is switched on, as well
as after on adjustable operating periode
between 5 and 120 minutes. The filter
cleaning process fakes approx. 6 seconds.
Continuous lighting of the red LED “Filter“
operating mode indication
signals “Filter
full“. On the next occasion when the suc-
tion motor is switched off, the filter will be
automatically cleaned for about 60 seconds.
This display also prompts the user to switch
off the connected unit briefly.
Flashing of the LED
indicates that the
filter is being cleaned and a filling level
measurement is taking place simultaneously.
A light up of the "Filter" operating mode
together wit the LED operating faults
indicates a fault of the filter or suction lines
(see operating faults 11.4 Error 24-filter
condition
1200)
.