background image

30

ABSORmatic

®

plus.

Full system serviceability only when
using original “KaVo filter car-

tridges“.

It should be ensured that suitable pro-
tective equipment (e.g. mouth protec-

tion, gloves, etc.) is worn.

When after a long period of use the filter
becomes worn to some extent, and this can
no longer be remedied by frequent “filter
cleaning“, the filter 

including the brush

strip 

must be changed. 

Take off the front panel 

and remove the

waste container (as described in 10.3)

Release the clamping nut 

on the filter

flange 

 

. Lift the filter flange 

 

slightly

and remove. 

Raise defective or damaged filter 

at the

front and withdraw it. Un-screw the used
brush strip 

. Fit new brush strip 

with

three knurled nuts. 
Pack used filters 

and brush strip 

in

PE waste bag (

Mat.- No. 0.228.3076

) and

place in waste disposal. 

Carefully insert new filter 

and align it

centrally in the rear filter flange.

Raise filter 

slightly and refix the filter

with the front filter flange 

 

by screw-

ing in the clamping nut 

Carry out the further assembly as described
in 10.3.

Switch off power switch 

and wait

approx. 20 sec., start again and the unit
returns to  normal operating mode.

ABSORmatic®

ABSORm

atic

®

ABSORmatic®

 

#

Содержание ABSORmatic plus

Страница 1: ...Operating Instructions ABSORmatic plus Always on the safe side...

Страница 2: ...KaVo Elektrotechnisches Werk GmbH Wangener Stra e 78 D 88299 Leutkirch Tel 0 75 61 86 150 Fax 0 75 61 86 265...

Страница 3: ...the load 10 A 5 4 Connection of external control display remote control connection 11 A 5 5 Connection of foot switch 11 A 6 Preparation for commencing operation 11 A 7 Control and functional element...

Страница 4: ...ormation improper use improperly performed repairs Repair and maintenance work apart from the activities described in these instructions for use may be performed only by quali fied technical personnel...

Страница 5: ...eaning and care of the unit In the case of newly painted surfaces on which water still forms droplets cleaning with water and nonabrasive mild cleaning agents is sufficient Paints subjected to stress...

Страница 6: ...d 42 mm Mat No 0 657 1182 1 x Tube bend 50 mm Mat No 0 657 1172 Operating instructions Mat No 1 000 2437 Connection parts for pressure line 1 x Mat No 0 220 0949 1 x Mat No 0 220 0455 1 x Mat No 0 203...

Страница 7: ...Suction hose 42 mm 1 2 m length Suction panel with control display suitable for EWL support frame Mat No 0 657 1531 incl Suction hose 42 mm 1 2 m length Exhaust air tubing set Mat No 0 657 0901 Consum...

Страница 8: ...sure of min 1 to max 6 bar The output operating pressure to the suction unit diverter is connected via a con trol unit with pressure regulating valve output approx 0 8 bar Correct use requires complia...

Страница 9: ...vered by the work bench the height of the dust collector must be adjusted Secure support panel with 4 screws and if necessary fix it with bench top clamps to the workplace bench top Place the bench co...

Страница 10: ...If necessary cut the suction hose to length so that it can be connected without hanging down also without being under tension Fit 50 mm suction hose onto pipe bend and fit pipe bend with suction hose...

Страница 11: ...ed into the suction unit without hanging down and also without being under tension Fasten suction hose to extraction point For connection to large equipment suction hose 50 mm fit straight socket and...

Страница 12: ...ad Connect the power cord of the suction route switching modul e g K9 ins talla tion K11 installation devesting and blast ing system etc With appropriately prepared units e g K Control controller the...

Страница 13: ...on and socket outlet for the dust collector A 5 5 Connection of foot switch Connect foot switch by means of snap plug to suction route switching modul ensure correct assignment A 6 Preparation for com...

Страница 14: ...operating mode Filter LED operating mode Waste container LED Functional disturbance Pushbutton Automatic mode Continuous mode Pushbutton Suction stage Symbol for filter Symbol for waste container Symb...

Страница 15: ...peration Pressing pushbutton switches from Continuous mode to Automatic mode In automatic mode suction only operates when at least one of the installations con nected e g K9 installation K11 installa...

Страница 16: ...8 3 Suction stage preselection Pushbutton allows manual switching from suction stages corresponding LED lights Programmed presetting of the suction air flow per suction point is obtained as follows A...

Страница 17: ...set and operated only in combination with the expansion set Errors warnings NO ERROR 00 ERROR 01 M2_ERROR 10 B8 CONTAINER ERROR 11 EEPROM_ERROR 23 ERROR DEFECTIVE 24 REDUCED PRESSURE_ERROR 25 NODEGUAR...

Страница 18: ...EEPROM basic setting yes 2251 ended no 2300 9 9 Automatic gain setting 2310 WP1 2301 Motor off 2311 WP2 2302 Motor on 2312 WP3 2303 Accept 2313 WP21 2304 Saved 2314 WP22 2305 Discarded 2315 WP23 2306...

Страница 19: ...up several times until display shows user program Press Enter Press pushbutton down or up and dial the requested utility program 9 2 Utility program User language Actuating pushbutton up or down to se...

Страница 20: ...Enter Select yes or no with pushbutton up or down Optimum operating mode yes This workplace obtains the selected suction power Other workplaces will be blocked in case there is not enough suction pow...

Страница 21: ...tinious operation Store selection with puhbutton Enter Factory setting Identification Current Actuating pushbutton up or down to select operating mode unit adap tion and press pushbutton Enter Select...

Страница 22: ...ptimum Yes Identification Current If the Modelgrinder is into operation WP1 automatically Sandblaster WP2 and Technician Workplace WP3 are blocked If Sandblaster WP2 is into operation the Technician W...

Страница 23: ...ear after the suction has been switched on Combination screen The requested suction power for the relevant workplace will be displayed Low pressure The actual low pressure will be displayed With pushb...

Страница 24: ...P3 by pressing foil switch up or down The machine to be measured has to be in stand by mode or turned off Press foil switch enter On the upper line of the display one can read motor off The lower para...

Страница 25: ...rrent 2400 By pressing foil switch up or down call up assistprogramm socket current and press foil switch enter Select the desirede work station WP1 WP2 WP3 by pressing foil switch up or down Press fo...

Страница 26: ...By pressing foil switch up or down call up assist programm carbon filt age and press enter The age of the active carbon filter is shown in worked hours With the foil switch left back you get out of t...

Страница 27: ...e cut off Once the memorised mem ory time lapse has elapsed the absormatic plus will start again when activated with the suchtion strength that was programmed in under the assist programm adjustment W...

Страница 28: ...container Lay the protruding material in the double edge of the container Press the waste bag against the contours of the waste container to keep trapped air to a minimum Place cardboard insert acces...

Страница 29: ...new active carbon blocks qty 2 so that the seal is towards the outer side Fit the hood bevel towards the front press on and lock with the two snap locks Carry out further installation as described in...

Страница 30: ...ating faults 11 7 Warning 30 Filter cleaning 1201 The filter is dirty and will be cleaned of the next operational interruption A short filter cleaning with filling level defection fakes place automati...

Страница 31: ...aste bag in the cleaned waste container and press it against the base and sides of the container Lay the protruding material in the double edge of the container Press the waste bag against the contour...

Страница 32: ...ng nut on the filter flange Lift the filter flange slightly and remove Raise defective or damaged filter at the front and withdraw it Un screw the used brush strip Fit new brush strip with three knurl...

Страница 33: ...2 Error message Error 11 1200 Dust container I Dust container has been pulled out or microswitch has malfunction C a Dust container has not been pulled back after cleaning b Micro switch defect or di...

Страница 34: ...ll suction hose or make proper connection c Install Filter lid 11 6 Error message Error 28 1200 EPROM defect I Component EPROM defect C The EPROM is tested if the unit is swit ched on through power sw...

Страница 35: ...id fine gold filter blocked R Clean dirt grid and possible exchange fine gold filter 11 11 Error message I Suction present at all 3 suction points although only one suction passage needed h C No compr...

Страница 36: ...Rated voltages Voltage ranges 100 V 50 60 Hz 120 V 60 Hz 230 V 50 60 Hz Rated power Suction max 1100 Watt for processing equipment connected max permissible 1900 Watt Connected load Rated input power...

Страница 37: ...4 m2 Max reduced pressure 180 mbar Suction power Stage I approx 60 86 unit Stage II approx 90 129 unit Stage III approx 135 167 unit Stage IV approx 141 191 unit 1 unit 0 9 1 0 m3 h Noise emission va...

Страница 38: ...ditional packaging Information on transport and unpacking is printed on the outside It applies to transport and storage Ambient conditions Permitted indors Permitted ambient temperature range of 14 F...

Страница 39: ...In order to make any claims for the damage KaVo requires the shipping note with confirmation of damage On no account return damaged goods to KaVo before contacting KaVo B 1 3 Pack weight and marking...

Страница 40: ...e parts or carry out repairs free of charge KaVo EWL accepts no liability for defects and their consequences which have arisen or could have arisen as a result of natural wear improper handling cleani...

Страница 41: ...tlige beskyttel seskrav i bestemmelserne i de f lgende direktiver johon t ss ilmoituksessa viitataan on direktiivien m r ysten t rkeimpien turvallisuusvaatimuksien mukainen som denne erkl ringen gjeld...

Страница 42: ...0 657 1382 0 657 0521 0 657 1142 0 657 0932 0 657 1232 0 245 5026 0 229 3005 0 221 1237 0 223 2232 0 696 0281 AT 1 000 3134 0 223 2733 T 5 A 0 266 4085 T 6 3 A 0 223 2813 T 3 15 A 0 657 1152 1 001 505...

Страница 43: ...19 100 120 V 0 223 2716 230 V 0 658 8732 42 mm 0 657 1472 2x 0 222 5123 0 657 1502 0 657 1482 AT 0 999 6206 0 695 1402 1 2 mtr 42 mm 0 695 1422 5 mtr 50 mm 0 657 0801 2x 0 657 1182 42 1x 0 657 1172 50...

Страница 44: ...01 GB 02 20 D D 8 88 82 29 99 9 L LE EU UT TK KI IR RC CH H T Te el le ef fo on n 0 07 75 56 61 1 8 86 6 1 15 50 0 F Fa ax x 0 07 75 56 61 1 8 86 6 2 26 65 5 I In nt te er rn ne et t w ww ww w k ka a...

Отзывы: