Visiomed - Kaufman KM-25 - V.2.3 -201012
15/21
IX.
T
ECHNICAL
&
P
RECISION INFORMATION
1.
Normal operating conditions :
Operating temperature:
Body temperature: 10°C ῀ 40°C
Surface temperature: 0°C to 50°C
Humidity : 10~93% RH
2.
Power : DC 3V (2 AAA batteries)
3.
Dimensions : 170 x 57 x 37 mm (L x W x H)
4.
Weight : 105g (without batteries)
5.
Display resolution : 0.1°C
6.
Measurement range :
Body mode : 32.9°C - 42.9°C
Surface mode : 0°C - 100.0°C
Room mode : 0°C - 100.0°C
7.
Consumption : ≤ 50mW
8.
Precision :
Body mode:
In a normal environment: ≤ ±0.3°C
Surface mode:
From 22 to 42°C: ≤ ±0.3°C
Above these temperatures ≤ ±2°C
Room mode:
From 10 to 40°C : ≤ ± 0.5°C
above 40°C : ≤ ±2°C
9.
Time taken per measurement : ≤ 0.8s
10.
Time for repeat measurement : ≤ ±0.2s
11.
Length of infrared waves : 5-14µm
12.
Measurement distance : 5 cm – 10 cm
13.
Automatic power off: 20 seconds
.
PRECISION OF THE KM-25
The Kaufman KM-25 thermometer can take
temperature readings below 32.9°C or above 42.9°C.
Precision is not guaranteed for readings in these
temperature ranges.
LONGEVITY OF USE
The Kaufman KM-25 thermometer is guaranteed to
last for 40,000 measurements.