![KAP Medical SKINGUARD Series Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/kap-medical/skinguard-series/skinguard-series_owners-manual_3425346008.webp)
SkinGuard® Series
Owner’s Manual
PN 400901 Rev. C
ECO 18-0281
Page 7
TECHNICAL SPECIFICATIONS
ELECTRICAL SPECIFICATIONS
Note: The control unit power inlet is used as the means of isolating the equipment from the supply mains on all
poles simultaneously.
U.S. / INTL.
Input Voltage AC:
90 ~ 264 VAC
Input Frequency:
60 / 50 Hz
Maximum Power Consumption:
SkinGuard
®
& SkinGuard
®
Turn
180 W
30 W
SkinGuard Float
®
40 W
30 W
Circuit Protection:
SkinGuard
®
& SkinGuard
®
Turn
Dual fused, 250V, 5A Fast blow fuse(s), std. fuses.
SkinGuard Float
®
Dual fused, 250V, 1A Slow blow fuse(s), std. fuses.
Fuse Type:
SkinGuard
®
& SkinGuard
®
Turn
Bussmann S500-5-R
SkinGuard Float
®
Bussmann GMD-1-R
Breaking Capacity: (BRK. CAP.)
SkinGuard
®
& SkinGuard
®
Turn
(BRK. CAP.) @125 VAC is 10kA
@250 VAC is 200A
SkinGuard Float
®
(BRK. CAP.) @125 VAC is 10kA
@250 VAC is 35A
Mode Of Operation:
Continuous
PERFORMANCE SPECIFICATIONS
Weight Capacity:
Standard Mattress: 500 Lb. (226 Kg.) maximum.
Bariatric Mattress: 1000 Lb. (455 Kg.) maximum.
Mattress Overlay: 250 Lb. (113 kg.) maximum.
U.S. / INTL.
Pressure Zones:
5
AP Zones:
2
Rotation Zones:
2
Max Flow:
SkinGuard
®
& SkinGuard
®
Turn
1275 LPM (45 CFM)
SkinGuard Float
®
42 ~ 52 LPM
Max Pressure:
35 ± 5 mmHg
Max Inflate Timer:
15 minutes
Support Surface Inflation Time:
SkinGuard
®
& SkinGuard
®
Turn
Within 60 seconds
SkinGuard Float
®
5 ~ 15 minutes
Patient Comfort Control Pressures
Soft Pressure:
6
5 mmHg
Firm Pressure:
SkinGuard
®
& SkinGuard
®
Turn
32
5 mmHg
SkinGuard Float
®
31
5 mmHg
AP Cycle Time:
5, 10, 15, 20 min
Turn Time:
1-253 min
Turn Angle:
5-45 degrees