SkinGuard® Series
Owner’s Manual
PN 400901 Rev. C
ECO 18-0281
Page 31
o
Press the TURN button to adjust the degree of rotation (angle). Degree of turn range is between
5-45 degrees. Use the ADJUST SETTINGS arrows to adjust the degree of rotation angle for the
Left side, then press the MAX INFLATE button to adjust the Right side.
o
Press TURN button to active the TURN mode based on the selected settings.
PAUSE
•
Press the PAUSE button to pause the turn cycle. Press it again to continue the turn cycle.
LOCK OUT
•
Control unit functions (including power) can be completely locked, preventing tampering of settings by
pressing and holding the LOCK button until the light comes on (approximately 3~5 seconds). Press and
hold the LOCK button to unlock.
ALARM SILENCE
•
An audio-visual alarm is sounded in the event of power failure or when the hose is disconnected from the
unit. Audio alarm can be muted by pressing alarm silence key.
POWER FAIL
•
In the event of a power outage, the microprocessor will activate an audiovisual signal to alert the
caregiver by flashing the amber “POWER FAIL” LED and the buzzer will turn on. Once the power is
restored to the control unit, the audiovisual signal will cease and the unit will resume operating its set
mode. During power outage the mattress will retain air as long as the mattress is connected to the control
unit.
LOW PRESSURE
“L” “P” flashing on display panel
•
In the event of hose disconnection, the microprocessor will activate an audiovisual signal to alert the
caregiver either by flashing “L” “P” on the display panel and turning on the buzzer or displaying Low
Pressure on the screen, flashing the amber “POWER FAIL” LED, and turning on the buzzer. Once the
low pressure problem is stable, the audiovisual signal will cease and the unit will resume operating its set
mode.
CPR FUNCTION
•
To deflate the mattress / overlay pad or for a CPR procedure, disconnect the mattress hose from the
control unit by pulling the mattress hose connector straight out from the control unit magnetic connector.
The CPR Valve on the mattress can also be used for a CPR procedure by rotating the CPR Valve to the
open position.