SkinGuard® Series
Owner’s Manual
PN 400901 Rev. C
ECO 18-0281
Page 17
SkinGuard® Turn Quick Reference Guide
SYMBOL
DESCRIPITON
EXPLANATION
POWER
Press the POWER button to turn on/off the control unit. The Standby (amber)
LED will turn off and the Operate (green) LED will turn on.
ADJUST
SETTINGS
Press the ADJUST SETTINGS Up/Down arrows to adjust the menu settings.
Follow specific menu instructions.
SOFTER /
FIRMER
Press the SOFTER/FIRMER arrows to adjust the patient comfort pressure
levels.
PATIENT
SETUP
Press the PATIENT SETUP button to set patient’s height and weight.
MAX INFLATE
Press the MAX INFLATE button to activate Max Inflate Mode. In MAX INFLATE
mode, the mattress will rapidly inflate. Beeps will occur every 3 minutes as a
reminder Max Inflate is active. Max Inflate will deactivate after 15 minutes.
SELECT
PRESS the SELECT button to choose between Therapy, Pulse, Multi, and
Fowler (Upright) modes.
TURN
Press the TURN button to select Turn Mode. Follow the instructions on the right
side of the screen to adjust the Left Turn, Right Turn, Dwell time, Left Degree,
and Right Degree of turn settings.
PAUSE
Press the PAUSE button to pause the turn cycle therapy time. Press PAUSE
again to reactivate turn cycle time.
ALARM
SILENCE
Press the ALARM SILENCE button (LED on) to mute the audio alarm.
LOCK
Press and hold the LOCK button (LED on) and all control unit functions will be
locked to prevent tampering of patient settings. Press and hold LOCK button to
turn off (LED off).
POWER FAIL
In the event of Power Failure, an audio/visual alarm will sound and display (LED
on).
LOW
PRESSURE
In the event of Low Pressure or if the hose is disconnected, an audio/visual
alarm will sound and display LOW PRESSURE in the window until correct
pressure is re-established.