K-1 & K-2 Series Owner’s Manual
Page 5
TECHNICAL SPECIFICATIONS
Note: The control unit Power Inlet is used as the means of isolating the equipment from the supply mains on all poles
simultaneously.
Electrical Specifications:
Input Voltage AC:
90 ~ 240 VAC
U.S. / INTL.
Input Frequency:
60 / 50 Hz
Maximum Power Consumption: 40 W
±
15 W
Circuit Protection:
Dual fused, 250V, 1A Slow blow fuse(s), std. fuses.
Fuse Type:
Bussmann GMD-1-R
Breaking Capacity: (BRK.CAP.) @125 VAC is 10kA
@250 VAC is 35A
Mode of Operation:
Continuous
Performance Specifications:
Mattress Overlay:
360 lbs. (163 Kg.) maximum.
Weight Capacity:
Standard Replacement Mattress:
360 lbs. (163 Kg.) maximum.
Bariatric Replacement Mattress:
1000 lbs. (455 Kg.) maximum.
Raised Side Bolstered Mattress:
360 lbs. (163 Kg.) maximum.
Bariatric Raised Side Bolstered Mattress:
1000 lbs. (455 Kg.) maximum.
Foam Aire Mattress:
360 lbs. (163 Kg.) maximum.
Pressure Zones:
2
AP Zones:
2
Max Flow:
42~52 LPM
Support Surface Inflation Time: 5~15 Minutes
Patient Comfort Pressures / Alternating Pressure
Soft Pressure:
6
±
5 mmHg
(K-1, K-2)
Firm Pressure:
32
±
5 mmHg (K-1) & 31
±
5 mmHg (K-2) of high AP pressure setting
A/P Time:
1 Minute - 99 Minutes
A/P Low Pressure:
0% - 80% (K-1 Elite) & 0% - 75% (K-2 Elite)
Patient Comfort Pressures / Alternating Pressure
Soft Pressure:
7
±
5 mmHg
(K-1oem, K-2oem)
Firm Pressure:
32
±
5 mmHg
A/P Time:
5, 10, 15 and 20 minute A/P Time(s)
A/P Low Pressure:
0%, 25% 50% or 75% (K-1oem) & 0% or 50% (K-2oem) of high AP pressure setting
Control unit and the mattress are
Patient Contact
lead free
,
mercury free
, and
latex free
Dartex top sheet is Halogen-free (bromide-free).
.
Mechanical Specifications: Control Unit (A)
Dimensions, LxWxH:
12” x 6 3/4” x 10.5” (30cm x 17cm x 27cm) (K-1, K-2 Elite models)
13.5” x 6” x 11” (34cm x 15cm x 28cm) (K-1oem, K-2oem models)
Weight:
K-1, K-1oem = 10.5 lbs. (4.8 Kg.), K-2 = 12 lbs. (5.5 Kg.) & K-2oem = 11 lbs. (5 Kg.)
Power Cord:
14’ long Hospital Grade
Connection:
(2) ¼” flow quick disconnect connector (K-1, K-1oem models)
(3) ¼” flow quick disconnect connector (K-2, K-2oem model)
Packaging:
1 piece per box
Air Filter
Internal (not serviceable)