Adjusting AP low pressure settings (K-1 and K-2 models) 

 

To adjust the low pressure zone of the Alternating Pressure therapy, follow the next steps: 
 

 

(K-1 Elite model) During Alternating Pressure therapy, press and hold (about 3 seconds) the (AP) AP key; A/P 
settings will be displayed on the screen. Use the (L) “+” and “-“ keys to adjust the Low AP percentage value (0% to 
80% mmHg of high pressure setting) . Press the AP key to save setting and continue AP therapy. Unit will time out 
and setting will be set automatically within 30 seconds. 
 
 

 

(K-2 Elite model) During Alternating Pressure therapy, press the (M) Mode key; A/P settings will be displayed on the 
screen. Use the (L) “+” and “-“ knob to adjust the Low AP percentage value (0% to 75% mmHg of high pressure 
setting), press AP key to save setting and continue AP therapy. Max can be press to activate ZIP (Zero Interface 
Pressure, 0% mmHg of high pressure setting). Use mode to switch therapy mode from the AP settings menu. Unit 
will time out and setting will be set automatically within 15 seconds. 

 

 

In AP mode the odd- numbered air cushions in the mattress will be maintained at a constant set patient comfort 
pressure and the even numbered air cushions deflate to the set percentage of the patient comfort high pressure in the 
first half of the AP cycle and vice versa for the second half of the cycle, and continue back and forth. 
 

PULSE (N) (WAVE)  
 

 

(K-2 Elite model) This mode is selected by pressing the (M) Mode key until “PULSE” is displayed on the screen.   
 

 

During Pulse mode, the pressure in each air cell will be decreased by 50% for 5 seconds and then increase to the prior 
pressure setting. This cycle will continue based on the selected Pulsation interval time. Using the (L) “+” and “-” knob 
adjust the Pulsation interval time from 1 to 99 minutes. 

 

 

MASSAGE  
 

 

(K-2 Elite model) This mode is selected by pressing (M) Mode key

 

until “MASSAGE” is displayed on the screen.   

 

 

In this mode the patient receives rapid alternating pressure (rapid high pressures and low pressures). 

 

 

MULTI-THERAPY 
 

 

(K-2 Elite model) To set Multi therapy mode, press Mode key (M)

 

until “MULTI” is displayed on the screen.  Use the TIME 

ADJUST knob (L) “+/-” to set the desired “Multi” therapy time. 

 

 

Multi-therapy is the combination of Static, A/P, Pulse and Massage therapies in K-2 Elite model. 
 
 

MODE KEY (M)  

 or 

 

 

 

The Mode key is used to cycle between therapy modes in K-1oem, K-2 Elite and K-2oem models. In addition to cycling 
between therapy modes, in K-2 Elite model the Mode key is also  used to configure the Patient Set-Up function and 
access the Alternating Pressure settings. 

 

 
 
 
 
 
 
 
 

Содержание K-1 Series

Страница 1: ...SYSTEM K 2 ELITE KARE ZONE DIGITAL ALTERNATING PRESSURE WITH ON DEMAND LOW AIR LOSS SYSTEM K 2oem ALT AIRE ALTERNATING PRESSURE WITH ON DEMAND LOW AIR LOSS SYSTEM US INTERNATIONAL PATENTS PENDING Ple...

Страница 2: ...ice 1 866 KAP MED 1 866 527 6331 Technical Support Service 1 951 340 4366 Email sales kapmedical com service kapmedical com www kapmedical com FDA Registered Company California FDA Registered Company...

Страница 3: ...ail Entrapment Risk Notification 9 Unit Overview 11 Explanation of Symbols Used 13 Mattress Overview 15 Unpacking the System 16 Mattress System Set Up 17 Operating Instructions 18 CPR Function 23 Pati...

Страница 4: ...pport combustion Failure to observe this warning can result in severe fire property damage and cause injury or death Entrapment may occur Patient entrapment with bed side rails and mattress may cause...

Страница 5: ...harmful interference received including interference that may cause undesired operation Changes or modifications not expressly approved by the party responsible for compliance could void the user s a...

Страница 6: ...2 LPM Support Surface Inflation Time 5 15 Minutes Patient Comfort Pressures Alternating Pressure Soft Pressure 6 5 mmHg K 1 K 2 Firm Pressure 32 5 mmHg K 1 31 5 mmHg K 2 of high AP pressure setting A...

Страница 7: ...n height replacement mattresses available along with 3 or 5 mattress overlays Packaging 1 piece per box Patented Low Air Loss top sheet on K 2 and k 2oem models only Optional mattress with Kevlar line...

Страница 8: ...California117 Optional California TB 106 TB 129 16 CFR 1632 16 CFR 1633 BS 6607 CRIB 5 BS 597 1 BS 597 2 compliant mattresses lined with Kevlar fire barrier available Kevlar lining based on flammabil...

Страница 9: ...control unit is not blocked and kept away from any heat sources or radiators during the operation of the unit Care should be taken such that the power cord of the control unit is not pinched or has an...

Страница 10: ...patient is unique Only the patient s medical care provider is familiar with the patient s unique medical condition and needs Only the patient s medical care provider and or the dealer from whom you o...

Страница 11: ...ces TSS which support the patient on an air mattress or specialized foam layer may present an increased risk of entrapment for some patients The benefit of TSS product use must be weighed against the...

Страница 12: ...Thermoplastic 94V 0 fire retardant enclosure houses a compressor main PCB display panel a short circuit over voltage protection with single dual fuse a quick connection and disconnection CPR deflatio...

Страница 13: ...s Owner s Manual Page 12 Refer to the Explanation of Symbols Used chart on page 13 for a cross reference of Display Key descriptions Unit Display Labels K 1 DERMA KARE K 1oem ALT AIRE K 2 KARE ZONE K...

Страница 14: ...STATIC T or Selects Static Therapy mode LOW AIR LOSS Y or or Activates Low Air Loss mode ALTERNATING PRESSURE AP or or Selects A P mode AP TIME TT or Selects AP cycle time FOWLER U or Fowler Mode The...

Страница 15: ...luded with each model SYMBOL EXPLANATION No Sharp Objects Low Heat Setting Do Not Dry Clean Do Not Bleach No Open Flames Normal Cycle Do Not Iron Latex Free Bolstered Mattress Even Bolstered Mattress...

Страница 16: ...top sheet has a special multi chambered air distribution layers which administers low air loss relief directly underneath the patient Top layer made out of urethane coated 70 Denier nylon or Dartex o...

Страница 17: ...of moving bed components before placing a patient on the bed and that the mattress system is fully inflated Operate all bed frame motorized functions through their full range of motion to be certain...

Страница 18: ...mattress system care should be taken such that the mattress is placed directly on the bed frame Figure 1 B Page 16 Unroll the Replacement Mattress and place it on the bed frame BF There are ten nylon...

Страница 19: ...m initialization routine for a few seconds Once the routine is completed the display D will read KAP MEDICAL STANDBY K 1oem K 2oem During initial power up when power cord Q is plugged into the power s...

Страница 20: ...atient s body There should be a minimum of 3 to 4 finger width clearance between the bottom of the patient and the safety foam base on an overlay there is no safety foam base Repeat this procedure unt...

Страница 21: ...even numbered air cushions deflate to the set percentage of the patient comfort high pressure in the first half of the AP cycle and vice versa for the second half of the cycle and continue back and f...

Страница 22: ...is factory programmed to a 35 or 45 angle depending on the customer s requirements It uses a 3 6V Lithium battery that can be replaced using KAP Medical part number 100622 S When the battery is first...

Страница 23: ...Cushion K 140 OPTIONAL SAC Optional foot support air cushion can be used to support patient s feet and to provide zero interface pressures in the heel area FOAM AIRE MATTRESS AS A NON POWERED MATTRESS...

Страница 24: ...mattress for a CPR procedure pull the hose V from the control unit connector press connector latches then pull connector If the mattress is equipped with bolsters and or air pad unplug the correspond...

Страница 25: ...unit flange connector PLEASE NOTE Before using the mattress please make sure that the mattress deflate connector is re connected back into the mattress deflate valve PATIENT TRANSPORTATION 1 Air Matt...

Страница 26: ...ray the exterior of the top and the bottom enclosures power cord and the cord plug with the prepared disinfectant detergent solution Using a damp cloth wipe down the sprayed surface cleanly Note Do no...

Страница 27: ...ote The Foam Aire mattresses have a Kevlar fire retardant sleeve sock barrier inside the cover Care should be taken if removing the cover for cleaning Before attempting to clean the mattress remove th...

Страница 28: ...o power source 3 Check power source has power and turn it ON 4 Wait till the power source has power 5 Send control unit back to factory for repair C Control Unit not responding 1 Unit locks up 1 Unplu...

Страница 29: ...ressure setting or 7 75 mmHg of high comfort pressure setting Use the Soft Firm keys to toggle between AP Low modes To end routine and accept setting press Power key Routine will auto time out K 2 ELI...

Страница 30: ...ICAL be liable for any loss direct indirect incidental or special damages arising out of or in connection with the use of this product The control unit warranty does not cover normal maintenance such...

Страница 31: ...K 1 K 2 Series Owner s Manual Page 30 NOTES Unit S N __________________________________ Mattress S N _______________________________ Auto Fowler S N ____________________________...

Страница 32: ...nnected securely to the control unit Plug the hospital grade power cord provided with the unit into a three pronged hospital socket The display will read KAP MEDICAL STAND BY Press POWER key SB to gre...

Страница 33: ...tress hose assembly is connected securely to the control unit Plug the hospital grade power cord provided with the unit into a three pronged hospital socket The amber light will illuminate Press POWER...

Страница 34: ...o set the patients height and weight Patient comfort settings will be automatic when done with settings Adjust the SOFT FIRM knob K to bypass automatic patient comfort pressure settings Actual mattres...

Страница 35: ...e power cord provided with the unit into a three pronged hospital socket The amber light will illuminate Press POWER key SB to green light position and then press Select key S until MAX Inflate LED W...

Отзывы: