6. Data Output
27
6.5 Analog Output (Optional)
(1)
Data Update Interval :
0.1 seconds
(2)
Load Impedance:
5K
Ω
or higher
(3)
Output Current :
DC
0~1V
For analog output, select one setting from the output range listed in the table
below. The measurement value displayed on the Normal Measurement screen
(Measurement Mode) will be output. For the setting procedure, refer to “7.2
Other Function Settings” in page 29.
Output Range
Conversion Formula
(Voltage: V)
Air Velocity (U)
0 ~ 5 m/s
U = 5×V
m/s
0 ~ 10 m/s
U = 10×V
m/s
0 ~ 25 m/s
U = 25×V
m/s
0 ~ 50 m/s
U = 50×V
m/s
0 ~ 1000 FPM
U = 1000×V
FPM
0 ~ 2000 FPM
U = 2000×V
FPM
0 ~ 5000 FPM
U = 5000×V
FPM
0 ~ 9999 FPM
U = 9999×V
FPM
Air Temperature (T) -10 ~ 40
℃
T =
50×V-10
℃
0 ~ 50
℃
T =
50×V
℃
0 ~ 100
℃
T = 100×V
℃
14 ~ 104
o
F
T = 90×V+14
o
F
32 ~ 122
o
F
T = 90×V+32
o
F
32 ~ 212
o
F
T = 180×V+32
o
F
Pressure (P)
-5 ~ +5 kPa
P = 10×V -5
kPa
-2 ~ +2 kPa
P = 4×V -2
kPa
Output Range: The output is linear output with the minimum value of 0V and maximum value of 1V.
The output data will be output every 0.1 seconds.
The configured time constant will be in effect. To change the time constant, refer to “3.4 Changing Time Constant”
in page 13.
Mode
How to take in measurement data (Analog output)
Explanation
TC1
Take data 10 times in 1
second and output the
average value of those
data as an instantaneous
value every 0.1 second.
TC5
Output the average value
over 5 seconds every 0.1
seconds. Data shifts by
every 0.1 seconds.
TC10
Output the average value
over 10 seconds every 0.1
seconds. Data shifts by
every 0.1 seconds.
Analog Output Terminal
0 2.5 5.0 7.5 10 sec (Measuring Time)
Average over 5 sec
0 0.5 1.0 1.5 2.0 sec (Measuring Time)
Average over 1sec
0 5 10 15 20 sec (Measuring Time)
Average over 10 sec