3.Display Description and Operation Procedure
58
· PUMP SETTING
In the PUMP setting, you can decide if the
pump is ON all the time or not.
To have the pump running all the time,
check the box for “PUMP ALWAYS ON”.
Tap [OK] to save the settings and close the
window. (
Authentication required
)
Tap [CANCEL] to discard any changes.
· REGION
In REGIONAL SETTING you can set the date
display format to be used when saving files.
Tap [OK] to save the settings and close the
window. (
Authentication required
)
Tap [CANCEL] to discard the current
settings.
*When transferring data and other files
to
a
PC,
different
date/time
or
timestamp will be displayed if the
setting of time zone is different from
the one in the PC.
Having the pump always on eliminates the
PREPARATION time before sampling.
However, it increases the chances of the
vane
becoming
worn
and
the
motor
deteriorating as the shaft of the motor may
go off-center from constant use.
We recommend the pump not be left running
during general use to extend the life of
pump.