3.Display Description and Operation Procedure
44
3.6.5 Saving Data
If AUTO SAVE is not checked on the MODE setting window, data will not be saved to a file after
measuring. If you try to change the MODE or start another measurement when AUTO SAVE is not
checked, you will get the message shown below.
- Tap [YES] to save the data.
- Tap [NO] to discard the data.
If the electricity goes out during a measurement, data will be treated as below.
[SINGLE] and [CONTINUOUS] Mode
Data during a measurement will be discarded when the power goes OFF.
[INTERVAL] Mode
Data taken before electricity goes out remains stored in the instrument.
When restarting the instrument, the message shown below will be displayed.
-
Tap [YES] to save the data.
-
The calculation data will be created using the
data collected before the power went out.
-
Tap [NO] to discard the data.
[STATS] and [STANDARD] Mode
Data taken before electricity went out remains in the instrument. However, the data from the
location where a measurement was being performed when the power went out is not saved.
The measurement will be in a temporary state of suspension.
-> Refer to
3.6.6 In Case that Measurement is Interrupted
.