3.Display Description and Operation Procedure
28
3.6 START
3.6.1 SINGLE, CONTINUOUS, INTERVAL
Tap [START] on the main screen to start a
measurement using the settings configured
in
3.5 MODE
.
If “START DELAY” is NOT checked in
“MODE” setting:
Upon tapping [START], the pump will
begin running. After the pump has run for
10 seconds to stabilize, the measurement
will begin
If only “START DELAY” is checked in
“MODE” setting:
Upon tapping [START], the Delay Time
begins. During the last 10 seconds of that
Delay Time the pump will run to stabilize,
after which the measurement will begin.
If “START DELAY” and “Set time”
are checked in “MODE” setting:
Upon tapping [START], “Input START
TIME” dialog box will be displayed. Enter
the
time
when
you
want
to
start
measuring and tap [OK]. During the last
10 seconds of the [Time left] time the
pump will run to stabilize, after which the
measurement will begin.
Tap [CANCEL] to go back to the menu
screen.
If you fail to press [OK] after entering
“START TIME”, and the entered START
TIME arrives, the instrument will increase
the start time in 10 minute intervals until
you press [OK].
.
<Input START TIME dialog box>